Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsDefibrillator Machine Parts

Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121

Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121

  • Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121
  • Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121
  • Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121
  • Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121
  • Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121
Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121
Product Details:
Brand Name: Nihon Kohden
Model Number: TEC-7621C
Payment & Shipping Terms:
Minimum Order Quantity: 1 Piece
Price: USD
Packaging Details: Standard Export Packing
Delivery Time: 3-5 working days
Payment Terms: T/T, Western Union, Paypal
Contact Now
Detailed Product Description
Brand: Nihon Kohden Product Category: Defibrillation
Product Type: Spare Parts Product Name: Relay Unit With HV Drive Board
Machine Application: TEC-7621C Relay Unit P/N: HV-761V
HV Drive Board P/N: UR-0121 Condition: Used
Package: Carton Delivery Date: 3 - 5 Working Days
High Light:

HV-761V Cardiolife TEC-7621C Defibrillator

,

UR-0121 Cardiolife TEC-7621C Defibrillator

Relay unit HV-761V with HV Drive Board UR-0121

Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V with HV Drive Board UR-0121

 

Brief Description

Relay Unit HV-761V with HV Drive Board UR-0121 for Nihon Kohden Cardiolife TEC-7621C Defibrillator

Relay Unit HV-761V with HV Drive Board UR-0121 - This part is second-hand but has complete performance and meets the usage requirements. It has been fully inspected by engineers before shipment to ensure its availability. You can always contact us before placing an order to know more details about this Relay Unit HV-761V with HV Drive Board UR-0121.

 

 

► Specification

Brand Nihon Kohden
Product Category Defibrillation
Product Type Spare Parts
Product Name Relay Unit with HV Drive Board
Relay Unit P/N HV-761V
HV Drive Board P/N UR-0121
Machine Application TEC-7621C
Condition Used with Good Condition
Warranty 90 Days
Delivery Date 3 - 5 Working Days
Package Carton
Payment Method Western Union, Bank T/T, Paypal
Transpotation DHL, FedEx, UPS, TNT, EMS, By Air, etc.

 

 

Romoving the HV Capacitor and Relay Unit

1. Remove the battery pack, memory card and lower case.

2. Remove the AC/DC unit.

3. Remove the main chassis and motherboard.

4. Short-circuit the two terminals of the high-voltage capacitor with a short bar, which is connected to the ground terminal on the wall through a ground wire.

5. Remove the four M3 flat-head screws with spring washers that secure the high-voltage capacitor to the lower case, and then remove the two high-voltage capacitor retaining brackets.

6. Remove the caps from the CNP552 and CNP553 wires from the high voltage capacitor.

7. Remove the high voltage capacitor.

8. Remove the three wires (CNP555, CNP554 and CNP556 wires) from the high voltage inductor that connect the high voltage inductor to the relay unit.

9. Pull out the relay unit.

NOTE: High voltage capacitors can store lethal energy. Always discharge the capacitor before touching any high-voltage component of the instrument (high-voltage capacitor, high-voltage inductor, relay unit, paddle socket, or paddle).

 
 
More Details

During the use of Nihon Kohden Cardiolife TEC-7621C Defibrillator, if there is a problem with the Relay Unit HV-761V, an error code will be displayed. Next, take a look at what the different error codes mean:

1. A501: During standby mode, the HV capacitor has more than 1 J energy more than one continuous second

2. A512: When charging starts, the high-voltage capacitor energy does not reach 1J within 2 seconds.

3. A513: The energy did not reach the selected energy within the specified time.

4. A524: After charging, the capacitor energy drops to the specified value for each energy.

5. A527: After charging, the capacitor energy is approximately 15% higher than the selected energy.

6. A529: After charging, the actual charging capacity is different from the selected capacity.

7. A556: Internal discharge takes more than 20 seconds to complete.

8. A566: 20 seconds after adjusting the internal discharge, the voltage of the high-voltage capacitor did not reach the target value.

9. A585: The high voltage capacitor voltage exceeds the specified voltage.

10. A587: When using disposable pads, 12.5% or more of the charged energy remains in the high-voltage capacitor after 2 seconds of external discharge.

 
 

Detailed Photos

Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121 0

Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121 1

 

 

 

 

 

 

 

 

Thanks for your interest in Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V with HV Drive Board UR-0121, please feel free to contact me. Thank you for your time and wish you have a happy life.

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Jasmine

Tel: 86-15180047496

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.