China World Technology Medical Equipment Service Group
Win-win together!
Brand Name: | Med-tronic |
Model Number: | LIFEPAK 12 LP12 |
Minimum Order Quantity: | 1 |
---|---|
Delivery Time: | 3-5 working days |
Payment Terms: | T/T, Western Union,PayPal |
Product Name: | Med-tronic Philipysio-Control Defibrillator / Monitor Series | SN: | LIFEPAK 12 LP12 |
---|---|---|---|
Product Superiority: | Defibrillator | Applications Scenes: | Med-tronic Philipysio-Control Defibrillator |
Key Word: | Defibrillator | After-sale Service: | In The Warranty Period And Non-human Damage Can Go To The Sale Free Of Charge For New |
Product Status: | In Good Form | Company Tenet: | Original Assembled Real Product, Quality Guarantee, Deliver On Time, Preference Price |
Stock Status: | Stock Available | Diagnostic:: | 0.05 To 150Hz Or 0.05 To 40Hz |
Monitor:: | 0.67 To 40Hz Or 1 To 30Hz | Paddles:: | 2.5 To 30Hz |
Analog ECG Output:: | 0.67 To 32Hz | ||
High Light: | PN 3006228 Defibrillator Standard Paddles,Lifepak LP15 Defibrillator Standard Paddles |
Medtronic Physio-Control Lifepak 12 Lifepak 15 LP12 LP15 Defibrillator Standard Paddles PN 3006228
► Brief Introduction
External sterilizable paddles are specifically designed to be used for cardiac defibrillation in a sterile environment. These paddles can be used for defibrillation, monitoring, and synchronized cardioversion and connect directly to the LIFEPAK 12 defibrillator/monitor.
► Features
Rugged Design: The device has a robust construction that can withstand rough handling and environments.
Biphasic Technology: It uses a biphasic waveform to deliver defibrillation therapy, which improves outcomes and reduces side effects.
User-Friendly Interface: The device has an intuitive user interface with a large display and simple controls, making it easy for medical professionals to operate.
Manual Defibrillation: The standard paddles allow medical professionals to manually deliver electric shocks to restore a normal heart rhythm.
Non-Invasive Pacing: The Lifepak 12 offers non-invasive pacing to stimulate the heart in case of bradycardia or heart block.
► Standard (hard) Paddles User Test
a | Examine cable for cracking, damage, broken or bent parts or pins. | Replace standard paddles/cable. |
b | Examine for paddle surface pitting or presence of dried or wet gel. | Replace standard paddles/cable, or clean paddles. |
c | Select PADDLES lead. | |
d | On paddles, turn ENERGY SELECT dial to 10 joules. | If a PADDLES LEADS OFF or CONNECT CABLE message appears, replace therapy cable and repeat check. If problem continues, remove the defibrillator from use and contact a qualified service technician. |
e | With paddles in paddle wells, press CHARGE button on paddles. | |
f | Press only one discharge button and release. Confirm energy was not discharged. | If energy discharges with one button press, repeat as stated above. |
g | Press the other discharge button and release. Confirm energy was not discharged. | |
h | Confirm screen message after pressing both discharge buttons: ABNORMAL ENERGY DELIVERED (biphasic defibrillator) --or-- ENERGY NOT DELIVERED (monophasic defibrillator) |
If message does not appear, replace therapy cable and repeat check. If problem continues, remove the defibrillator from use and contact a qualified service technician. |
i | Remove paddles from wells, and confirm artifact on screen. | If task fails, replace therapy cable and repeat check. If problem continues, remove the defibrillator from use and contact a qualified service technician. |
j | Place paddle surfaces together, and confirm flat line on screen. |
► Specification
Energy Level (J) | Acceptable Output (J) |
2 | 1.0 to 3.0 |
10 | 9.0 to 11.0 |
70 | 65.1 to 74.9 |
200 | 186.0 to 214.0 |
360 | 334.8 to 385.2 |