Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsDefibrillator Machine Parts

PN 3006228 Defibrillator Standard Paddles Medtronic Physio Control Lifepak 12 Lifepak 15 LP12 LP15

PN 3006228 Defibrillator Standard Paddles Medtronic Physio Control Lifepak 12 Lifepak 15 LP12 LP15

  • PN 3006228 Defibrillator Standard Paddles Medtronic Physio Control Lifepak 12 Lifepak 15 LP12 LP15
  • PN 3006228 Defibrillator Standard Paddles Medtronic Physio Control Lifepak 12 Lifepak 15 LP12 LP15
  • PN 3006228 Defibrillator Standard Paddles Medtronic Physio Control Lifepak 12 Lifepak 15 LP12 LP15
  • PN 3006228 Defibrillator Standard Paddles Medtronic Physio Control Lifepak 12 Lifepak 15 LP12 LP15
PN 3006228 Defibrillator Standard Paddles Medtronic Physio Control Lifepak 12 Lifepak 15 LP12 LP15
Product Details:
Brand Name: Med-tronic
Model Number: LIFEPAK 12 LP12
Payment & Shipping Terms:
Minimum Order Quantity: 1
Delivery Time: 3-5 working days
Payment Terms: T/T, Western Union,PayPal
Contact Now
Detailed Product Description
Product Name: Med-tronic Philipysio-Control Defibrillator / Monitor Series SN: LIFEPAK 12 LP12
Product Superiority: Defibrillator Applications Scenes: Med-tronic Philipysio-Control Defibrillator
Key Word: Defibrillator After-sale Service: In The Warranty Period And Non-human Damage Can Go To The Sale Free Of Charge For New
Product Status: In Good Form Company Tenet: Original Assembled Real Product, Quality Guarantee, Deliver On Time, Preference Price
Stock Status: Stock Available Diagnostic:: 0.05 To 150Hz Or 0.05 To 40Hz
Monitor:: 0.67 To 40Hz Or 1 To 30Hz Paddles:: 2.5 To 30Hz
Analog ECG Output:: 0.67 To 32Hz
High Light:

PN 3006228 Defibrillator Standard Paddles

,

Lifepak LP15 Defibrillator Standard Paddles

Medtronic Physio-Control Lifepak 12 Lifepak 15 LP12 LP15 Defibrillator Standard Paddles PN 3006228

 

 

 

► Brief Introduction

External sterilizable paddles are specifically designed to be used for cardiac defibrillation in a sterile environment. These paddles can be used for defibrillation, monitoring, and synchronized cardioversion and connect directly to the LIFEPAK 12 defibrillator/monitor.

 

 

 

► Features

Rugged Design: The device has a robust construction that can withstand rough handling and environments.

Biphasic Technology: It uses a biphasic waveform to deliver defibrillation therapy, which improves outcomes and reduces side effects.

User-Friendly Interface: The device has an intuitive user interface with a large display and simple controls, making it easy for medical professionals to operate.

Manual Defibrillation: The standard paddles allow medical professionals to manually deliver electric shocks to restore a normal heart rhythm.

Non-Invasive Pacing: The Lifepak 12 offers non-invasive pacing to stimulate the heart in case of bradycardia or heart block.

 

 

 

Standard (hard) Paddles User Test

a Examine cable for cracking, damage, broken or bent parts or pins. Replace standard paddles/cable.
b Examine for paddle surface pitting or presence of dried or wet gel. Replace standard paddles/cable, or clean paddles.
c Select PADDLES lead.  
d On paddles, turn ENERGY SELECT dial to 10 joules. If a PADDLES LEADS OFF or CONNECT CABLE message appears, replace therapy cable and repeat check. If problem continues, remove the defibrillator from use and contact a qualified service technician.
e With paddles in paddle wells, press CHARGE button on paddles.
f Press only one discharge button and release. Confirm energy was not discharged. If energy discharges with one button press, repeat as stated above.
g Press the other discharge button and release. Confirm energy was not discharged.
h Confirm screen message after pressing both discharge buttons:
ABNORMAL ENERGY DELIVERED (biphasic defibrillator)
--or--
ENERGY NOT DELIVERED (monophasic defibrillator)
If message does not appear, replace therapy cable and repeat check. If
problem continues, remove the defibrillator from use and contact a qualified service technician.
i Remove paddles from wells, and confirm artifact on screen. If task fails, replace therapy cable and repeat check. If problem continues, remove the defibrillator from use and contact a qualified service technician.
j Place paddle surfaces together, and confirm flat line on screen.

 

 

 

► Specification

Energy Level (J) Acceptable Output (J)
2 1.0 to 3.0
10 9.0 to 11.0
70 65.1 to 74.9
200 186.0 to 214.0
360 334.8 to 385.2

 

PN 3006228 Defibrillator Standard Paddles Medtronic Physio Control Lifepak 12 Lifepak 15 LP12 LP15 0

 

 

 

Please kindly contact us, if you are interested in Medtronic Physio-Control Lifepak 12 Lifepak 15 LP12 LP15 Defibrillator Standard Paddles PN 3006228, Thank you!

 

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Rita

Tel: 8618279091952

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.