22406_Rev.005 XD100xe M290 Defibrillator Board Metrax Gmbh Green Color

22406_Rev.005 XD100xe M290 Defibrillator Board Metrax Gmbh Green Color

  • 22406_Rev.005 XD100xe M290 Defibrillator Board Metrax Gmbh Green Color
  • 22406_Rev.005 XD100xe M290 Defibrillator Board Metrax Gmbh Green Color
  • 22406_Rev.005 XD100xe M290 Defibrillator Board Metrax Gmbh Green Color
  • 22406_Rev.005 XD100xe M290 Defibrillator Board Metrax Gmbh Green Color
  • 22406_Rev.005 XD100xe M290 Defibrillator Board Metrax Gmbh Green Color
22406_Rev.005 XD100xe M290 Defibrillator Board Metrax Gmbh Green Color
Product Details:
Brand Name: Primedic
Model Number: 22406_Rev.005
Payment & Shipping Terms:
Minimum Order Quantity: 1pcs
Packaging Details: Carton
Delivery Time: 3-5 working days
Contact Now
Detailed Product Description
Payment Terms: Full Before Shipment Product Name: PRIMEDIC DefiMonitor XD100xe(M290) Defibrillator Board Metrax Gmbh 22406_Rev.005 XTBoard2.11 C8fC
Brand: Primedic Part No .: 22406_Rev.005
Use For: PRIMEDIC DefiMonitor XD100xe(M290) Defibrillator MOQ: One Piece
Warranty: 90 Days Condition: Used
Color: Green Packaging: Standard
High Light:

M290 Defibrillator Board

,

22406_Rev.005 Defibrillator Board

22406_Rev.005 XTBoard2.11 C8fC PRIMEDIC DefiMonitor XD100xe(M290) Defibrillator Board Metrax Gmbh 

 

 

 

 

PRIMEDIC DefiMonitor XD brief introduction

 

PRIMEDIC DefiMonitor XD is an external defibrillator with integrated 6-channel ECG.

ECG can be recorded via PRIMEDIC SavePad, defibrillation paddles or 3-pin patient cable with adhesive ECG electrodes. PRIMEDIC DefiMonitor XD is available in a variety of configurations. The appropriate model number will be displayed in the monitor's welcome dialog.

In automatic mode, the ECG is analyzed by an implemented algorithm. If a life-threatening cardiac arrhythmia is detected, the device alerts the user of defibrillation and, when activated by the user, delivers the shock necessary to restore resistance. If the device does not detect an abnormal rhythm that requires defibrillation, no shock is delivered.

 

 

 

XtBoard Error 

 

Error text Cause(s) User action
 XtBoard err - power source Error during the current supply of the expansion board

- charge and/or replace the power module

- if the error persists: Return the device for servicing

XtBoard err - paddle logic Error during the test of the paddles

- do not press any paddle buttons when starting the device

 

- remove and reinsert the paddles

 

- restart the device

 

- if the error persists: Return the device for servicing

XtBoard err - key logic Error during the test of the paddle buttons

- restart the device

- do not press any paddle buttons when starting the device

- if the error persists: Return the device for servicing

XtBoard err - unknown General fault message expansion board

- restart the device

-if the error persists: Return the device for servicing

XtBoard err - runtime error Error during the expansion board runtime check

- restart the device

 

-if the error persists: Return the device for servicing

XtBoard err - platform Error during the hardware detection -return the device for servicing

 

 

FAQ

1. Whats the shipping cost ?

The shipping cost depends on packaging weight and volume .It will be appreciated to confirm products’ quantity .Then we will help you check the shipping cost .If you have your own Chinese forwarder or DHL ,Fedex account ,then free shipping

 

2.Do you ship by plane or by Sea

Yes ,we do .We ship by DHL ,Fedex or UPS ,around 5-7 working days to arrive on your side .For battery ,we will ship by a special liner way ,which will take longer .If urgency ,pls arrange the order in advance .Thank you for your cooperation in advance .

 

3.How long is your warranty

Our warranty is 90 days .If you have any after-sales problem ,feel free to contact us .

 

22406_Rev.005 XD100xe M290 Defibrillator Board Metrax Gmbh Green Color 0

 

Thank you so much for your attention  .If you're interested in 22406_Rev.005 XTBoard2.PRIMEDIC DefiMonitor XD100xe(M290) Defibrillator Board Metrax Gmbh ,feel free to cotnact us .Thank you .

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Kiara

Tel: 8619854815217

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.