China World Technology Medical Equipment Service Group
Win-win together!
Brand Name: | Goldway |
Model Number: | GW5A1E02C |
Minimum Order Quantity: | 1 Piece |
---|---|
Price: | USD |
Packaging Details: | Standard Export Packing |
Delivery Time: | 3-5 working days |
Payment Terms: | T/T, Western Union, Paypal |
Brand: | Goldway | Product Category: | Monitoring |
---|---|---|---|
Product Type: | Spare Parts | Item Type: | Board |
Item Name: | ECG Board | Application: | UT4000F PRO |
Model: | GW5A1E02C | Condition: | Used |
Performance: | Good | Warranty: | 90 Days |
Delivery Date: | 3 - 5 Working Days | In Stock: | Yes |
High Light: | Philip Goldway UT4000F Patient Monitor Parts,GW5A1E02C Patient Monitor Parts |
ECG Board GW5A1E02C for Goldway UT4000F PRO
Philip Goldway UT4000F PRO Patient Monitor ECG Board GW5A1E02C Medical Equipment Spare Parts, Circuit Board
► Brief Description
ECG Board GW5A1E02C for Philip Goldway UT4000F PRO Patient Monitor, Spare Part for Medical Equipment
ECG Board GW5A1E02C for Goldway UT4000F PRO Monitor - There are many very important parts on the UT4000F PRO patient monitor, because the device can monitor a lot of body data for the patient, including ECG, SPO2, TEMP, NIBP, heart rate, pulse, etc. This ECG Board GW5A1E02C is the main circuit board that can support the normal operation of the ECG part of the device. This board can transmit the received ECG signal to the main board, and then the main board transmits it to the monitor screen, presenting it to the user in a more intuitive and fast form for reference.
We provide 90 days warranty to all of our products, please contact me freely is there have any problems.
► Specification
Brand: | Philip Goldway |
Product Category: | Monitoring |
Product Type: | Spare Parts |
Item Type: | Board |
Item Name: | ECG Board |
Item Number: | GW5A1E02C |
Machine Application: | UT4000F PRO |
Condition: | Working in Good Condition |
Warranty: | 90 Days |
Transpotation: | DHL, FedEx, UPS, By Air, etc. |
Material: | Plastic, Metal, etc. |
► More Details
The ECG Board GW5A1E02C is installed on the Philip Goldway UT4000F PRO Patient Monitor. But before performing ECG monitoring on your patient, be sure to perform an ECG test:
ECG TEST:
1. First, set the ECG simulator data to a heart rate of 80bpm and an ECG gain of 1mV;
2. Secondly, connect one end of the ECG Cable and lead wire to the ECG connector of the UT4000F monitor, and connect the other end to the simulator;
3. Make sure the analog device and monitor are connected correctly;
4. It is necessary to observe that the ECG channel shows no murmur in ECG lead II, and the heart rate shows 80±2bpm;
5. Check each ECG lead for ECG waveform and noise interference;
6. Adjust the ECG gain to 2mV and observe whether there is any change in the ECG waveform;
7. Unplug the RA from the simulator and observe whether the screen prompts "lead off";
8. Reconnect the RA lead and observe;
9. Set 1mV calibration in the setting menu and observe whether there is a 1mV calibration square wave overlapping on the ECG waveform to complete the test.