China World Technology Medical Equipment Service Group
Win-win together!
Brand Name: | Med-tronic |
Model Number: | 21330-001176 |
Minimum Order Quantity: | 1pcs |
---|---|
Packaging Details: | Carton |
Delivery Time: | 3-5 working days |
Payment Terms: | T/T(wire transfer), Western Union, Paypal |
Product Name: | REF21330-001176 Med-tronic PHiliphYSIO-CONTROL LIFEPAK 15 LP 15 LITHIUM-ION Rechargeable Battery | Brand: | Med-tronic |
---|---|---|---|
Part No .: | 21330-001176 | Battery Type: | Lithium-ion |
Weight: | 0.59 Kg (1.3 Lb) | Voltage: | 11.1V Typical |
Capacity (rated): | 5.7 Amp Hours | Charge Time (with Fully Depleted Battery): | 4 Hours And 15 Minutes (typical) |
Battery Indicators: | Each Battery Has A Fuel Gauge That Indicates Its Approximate Charge. A Fuel Gauge That Shows Two Or Fewer LEDs After A Charge Cycle Indicates That The Battery Should Be Replaced. | Charging Temperature Range: | 0° To 50°C (32° To 122°F) |
Operating Temperature Range: | 0° To 50°C (32° To 122°F) | Short Term (<1 Week) Storage Temperature Range: | -20° To 60°C (-4° To 140°F) |
Long Term (>1 Week) Storage Temperature Range: | 20° To 25°C (68° To 77°F) | Operating And Storage Humidity Range: | 5 To 95% Relative Humidity, Non-condensing |
Highlight: | REF21330-001176 Defibrillator Machine Parts,Lithium Ion Battery Defibrillator Machine Parts,Defibrillator Lithium Ion Rechargeable Battery |
REF21330-001176 Med-tronic PHiliphhilipYSIO-CONTROL LIFEPAK 15 LP 15 LITHIUM-ION Rechargeable Battery
Med-tronic lifepak 15 device is powered by two LIFEPAK 15 rechargeable lithium-ion (Li-Ion) batteries. The battery communicates through contacts on the bottom of the battery. This allows you to exchange of information about battery type, amp hours rating, charge rate, target voltage, current, and other parameters
Battery Specifications
CHARACTERISTIC | DESCRIPTION |
Battery Type | Lithium-ion |
Weight | 0.59 kg (1.3 lb) |
Voltage | 11.1V typical |
Capacity (rated) | 5.7 amp hours |
Charge Time (with fully depleted battery) | 4 hours and 15 minutes (typical) |
|
Each battery has a fuel gauge that indicates its approximate charge. A fuel gauge that shows two or fewer LEDs after a charge cycle indicates that the battery should be replaced. |
Charging Temperature Range | 0° to 50°C (32° to 122°F) |
Operating Temperature Range | 0° to 50°C (32° to 122°F) |
Short Term(<1 week) Storage Temperature Range | -20° to 60°C (-4° to 140°F) |
Long Term (>1 week) Storage Temperature Range | 20° to 25°C (68° to 77°F) |
Operating and Storage Humidity Range | 5 to 95% relative humidity, non-condensing |
The LIFEPAK 15 lithium-ion battery, charger and power cord cannot be replaced by the batteries, charger and power cord used in other LIFEPAK defibrillators.
The home screen shows a battery indicator that shows the following information about the battery installed in the device:
• With or without batteries in the battery compartment.
• The battery is in use.
• Battery charging mode.
When installing two batteries The device will use the battery with the lowest charge level first. Used batteries are indicated by the white battery number in the black box. when the battery reaches the mode "Replace battery", the device will automatically switch to a different battery.
Indicator | Meaning | Description |
![]() |
Active battery | The device is running on battery in Well 1. The battery status indicator shows up to four green bars. Each green bar represents approximately 25% remaining charge. For example, three green bars indicate about 75% remaining charge. |
![]() |
Low battery | The battery in Well 1 is in use and low. A yellow bar indicates 5% to 10% charge remaining. |
![]() |
Very low battery | The battery in Well 1 is in use and is very low. A flashing red bar indicates 0% to 5% battery remaining. The device will automatically switch to another battery only when the battery is fully charged. If both batteries show red bars, the battery replacement voice prompt appears. |
![]() |
Unrecognized battery | The battery in well 2 is not used. Battery communication failed or non-PHiliphhilipysioControl battery installed. The battery can power the device, but the charge level is unknown and no low battery messages or warnings are displayed |
![]() |
No battery installed or fault detected | No batteries are installed in well 1, or a battery error has been detected in well 1 and the tool is not using batteries. |
Battery Fuel Gauge
The LIFEPAK 15 Li-Ion battery has a push-button fuel gauge and is able to communicate with the Li-Ion battery charger.
The push-button fuel gauge provides a visual indication of battery capacity and battery condition through a series of four green LEDs. Press this button to light the LED in one of the following modes:
# Fuel Gauge LEDs | Absolute State of Charge (ASOC) | Messaging/Comments |
4 green | 75% < ASOC | None |
3 green | 50% < ASOC | None |
2 green | 25% < ASOC | None |
1 green | 15% < ASOC | None |
1 yellow | 5% < ASOC | LOW BATTERY |
1 flashing red | ASOC £ 5% | REPLACE BATTERY |
Battery Operation Only
Maximum Time from Charge to Shock Ready (Manual Mode)
Battery Status | Charge Time |
Fully charged | 200 J within 7 seconds, nominal |
Fully charged, followed by 15 full-energy shocks | 360 J within 10 seconds |
Fully charged | 360 J within 10 seconds |
Maximum Time from Initiation of Analysis to Shock Ready (AED Mode):
Battery Status | Charge Time |
Fully charged | 200 J within 15 seconds, nominal |
Fully charged, followed by 15 full-energy shocks | 360 J within 30 seconds |
Fully charged | 360 J within 30 seconds |
Maximum Time from Power-on to Shock Ready (Manual Mode):
Battery Status | Charge Time |
Fully charged, followed by 15 full-energy shocks | 360 J within 25 seconds |
Maximum Time from Power-on to Shock Ready (AED Mode):
Battery Status | Charge Time |
Fully charged, followed by 15 full-energy shocks | 360 J within 40 seconds |
If you’re interested in REF21330-001176 Med-tronic PHiliphhilipYSIO-CONTROL LIFEPAK 15 LP 15 LITHIUM-ION Rechargeable Battery ,feel free to contact us .Thank you .
Contact Person: Kiara