China World Technology Medical Equipment Service Group
Win-win together!
Brand Name: | Philip Goldway |
Model Number: | UT4000F |
Minimum Order Quantity: | 1pcs |
---|---|
Packaging Details: | Carton |
Delivery Time: | 3-5 working days |
Payment Terms: | T/T(wire transfer), Western Union, Paypal |
Product Name: | Used Philip Goldway UT4000F Patient Bedside Monitor | Brand: | Philip Goldway |
---|---|---|---|
Model No: | UT4000F | MOQ: | One Piece |
Warranty: | 90 Days | Condition: | Used |
Power Supply: | AC 22V±22V 50Hz | Parameter: | SPO2 ,ECG ,TEMP,NIBP |
Display: | 10.4" TFT Color Display | Packaging: | Carton Box |
High Light: | 10.4'' TFT Display Used Patient Monitor,Philip Goldway UT4000F Used Patient Monitor,Multi Parameter Patient Bedside Monitor |
Used Philip Goldway UT4000F Multi-Parameter Patient Bedside Monitor
Philip Goldway UT4000F Patient Monitor
About Philip Goldway UT4000F Patient Monitor
The Philip Goldway UT4000F patient monitor is a medical device used to monitor patients' vital signs and parameters in a healthcare setting.
Some of the key features of the UT4000F are:
Troubleshooting for Philip Goldway UT4000F Patient Monitor
► when the instrument is turned on. There is no on-screen display and no indicator light.
Monitoring and analysis: If the instrument is connected to AC power without being turned on, the power light should be orange (or red); after switching on. The light should be green. Therefore:
(1) Check whether the mains socket and the two fuses on the back of the instrument are normal, that is, check whether there is AC input;
(2) Turn on the instrument and determine whether the switching power supply has a 15V DC output if there is a 220V AC input;
(3) If the instrument is turned on without AC power, the display will not appear and the indicator will not light. Indicates that the battery is low or damaged.
Solution:
(1) Connect all connecting parts reliably and make sure there is AC input;
(2) If your switching power supply does not have 15V output, you need to replace the switching power supply:
(3) Connect AC power to charge the instrument. If the battery is still low after 24 hours of charging, it means the battery is damaged and needs to be replaced.
► When the meter is turned on, the screen is blank. The light is always orange (or red).
Control and analysis: The light is orange. Indicates that the switching power supply has a 15V DC output:
(1) Turn on the instrument and use a multimeter to check whether the power switch is damaged;
(2) Check that the connection cable between the switch and the motherboard is not loose or damaged.
Solution:
(1) Replace the power switch;
(2) Connect the cable reliably; if it does not turn on, replace the motherboard.
► When the meter is turned on, there is no display on the screen, but the light is green.
Check and Analysis: If the button can start inflation at this time or you hear a heartbeat while measuring blood oxygen, it means the meter is turned on, but the TFT screen does not light up. Open the tool and check:
(1) There is an input voltage between 2.5V and 5.4V on the inverter power board or the connector is loose:
(2) Use a multimeter to check whether the inverter has high voltage output.
Solution:
(1) Connect the connection cable reliably and ensure the 2.5-5.4V output:
(2) If there is no high voltage output, replace the inverter.
► When the meter is turned on, the screen is blank. But there is a raster (commonly known as white screen) or the display colors are messed up.
Control and analysis: the TFT screen lights up. But the display part of the PC104 board is defective. The display signal cannot be output correctly
Solution: replace the PC104 board.
Specification of Philip Goldway UT4000F Patient Monitor
Product Name |
Philip Goldway UT4000F Patient Monitor
|
Brand | Philip Goldway |
Model no | UT4000F |
Condition | Used |
Power supply | AC 22V±22V 50Hz |
Parameter | SPO2 ,ECG ,TEMP,NIBP |
MOQ | One piece |
Warranty | 90 days |
Thank you so much for your attention .If you're interested in Used Philip Goldway UT4000F Multi-Parameter Patient Bedside Monitor ,feel free to contact us .Thank you .
Contact Person: Kiara
Tel: 8619854815217