Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Parts

Philip VS2+ Patient Monitor Parts Parameter NBP Spo2 Panel Board Connector 453564200151

Philip VS2+ Patient Monitor Parts Parameter NBP Spo2 Panel Board Connector 453564200151

  • Philip VS2+ Patient Monitor Parts Parameter NBP Spo2 Panel Board Connector 453564200151
  • Philip VS2+ Patient Monitor Parts Parameter NBP Spo2 Panel Board Connector 453564200151
  • Philip VS2+ Patient Monitor Parts Parameter NBP Spo2 Panel Board Connector 453564200151
  • Philip VS2+ Patient Monitor Parts Parameter NBP Spo2 Panel Board Connector 453564200151
  • Philip VS2+ Patient Monitor Parts Parameter NBP Spo2 Panel Board Connector 453564200151
  • Philip VS2+ Patient Monitor Parts Parameter NBP Spo2 Panel Board Connector 453564200151
Philip VS2+ Patient Monitor Parts Parameter NBP Spo2 Panel Board Connector 453564200151
Product Details:
Brand Name: Philip
Model Number: 453564200151
Contact Now
Detailed Product Description
Product Name: Philip VS2+ Patient Monitor Parameter NBP Spo2 Panel Board Connector 453564200151 Brand: Philip
Use For: Philip VS2+ Patient Monitor MOQ: One Piece
Warranty: 90 Days Condition: Used
Part No: 453564200151 Material: Plastic
High Light:

453564200151 Patient Monitor Parts

,

NBP Spo2 Panel Patient Monitor Parts

,

NBP Spo2 Panel Board Connector

Philip VS2+ Patient Monitor Parameter NBP Spo2 Panel Board Connector 453564200151

 

 Parameter Panel Board 453564200151 of  Philip VS2+ Patient Monitor is made of plastic ,which includes NBP and Spo2 connector .The panel connector is to connect NBP and Spo2 accessories to make NBP and Spo2 function 

 

 

NBP Test

 

The NBP test examines the performance of noninvasive blood pressure measurements. When checking the NBP module, perform the following steps:

 

  • NBP Accuracy NBP Calibration Procedure (if required)

 

  • NBP Pneumatic Leak Test

 

To perform this test, you need the following tests:

 

  • Reference pressure gauge (including hand pumps and valves)

 

  • Reading expansion chamber accuracy 0.2% (volume 500 ml ± 10%)

 

  • Tube

 

 

 

 

NBP problem

 

Symptoms Possible cause Action
The NBP cuff will not inflate. The tube or cuff is bent or twisted. Straighten the tube or cuff.
Air leaks from tube or cuff. Replace the cuff and make sure there are no air leaks in the tube.
The NBP module is malfunctioning.

Replace the NBP module.

replace the motherboard

NBP measurements are not displayed. Wrong size cuff or improper cuff placement. Use the correct cuff size and ensure correct cuff placement.
There is a problem external to the monitor. Make sure the tubing is straight and free of kinks. Replace the cuff and make sure there are no air leaks in the tube.
Incorrect initial inflation pressure setting. Ensure all external blood pressure reading requirements are met and the patient is not excessively mobile
An NBP module error has occurred.

Replace the NBP module.

replace the motherboard

NBP measurements are not reliable The cuff is the wrong size, or the cuff is placed incorrectly. Use the correct cuff size and ensure correct cuff placement.
There is an external problem. Ensure that all external blood pressure reading requirements are met and that the patient is not excessively mobile.
Incorrect initial inflation pressure setting. Specify the correct initial inflation pressure.
Calibration is not reliable. Recalibrate the NBP module
NBP initial pressure is too high. The cuffs are too tight. Make sure the cuff is properly fastened properly.
Incorrect initial inflation pressure setting. Specify the correct initial inflation pressure.

 

 

 

Removing the NBP module

 

Caution

 

After replacing the NBP module, the NBP cycle count must be reset and the NBP test run.

 

 

 

1.Separate the front panel from the rear case assembly.

 

2. Disconnect the NBP cable from the motherboard.

 

 

Philip VS2+ Patient Monitor Parts Parameter NBP Spo2 Panel Board Connector 453564200151 0

 

3.Disconnect the NBP tubing from the motherboard.

 

Philip VS2+ Patient Monitor Parts Parameter NBP Spo2 Panel Board Connector 453564200151 1

 

4.Remove the three Torx T10 screws from the metal bracket that attaches the NBP module assembly to the motherboard

 

 

Philip VS2+ Patient Monitor Parts Parameter NBP Spo2 Panel Board Connector 453564200151 2

 

5.Remove the assembly from the front housing.

 

 

Thank you for your attention .If you're interested in Philip VS2+ Patient Monitor Parameter NBP Spo2 Panel Board Connector 453564200151 ,feel free to contact us .Thank you .

 

 

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Kiara

Tel: 8619854815217

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.