Edan IM50 patient monitor parameter panel board brief introduction
Edan IM50 patient monitor parameter panel board is part of the patient monitoring system that displays various parameters related to the patient's vital signs. This panel typically displays information such as heart rate, blood pressure, oxygen saturation, respiratory rate, temperature, and ECG waveforms. It allows healthcare professionals to monitor a patient's condition in real time and intervene as necessary. Panels can also allow alerts to be set when certain values fall outside normal ranges, enabling healthcare professionals to quickly respond to potential problems. Overall, a patient monitor parameter panel is an essential part of a patient monitoring system that ensures continuous monitoring and maintenance of patient safety and health.
Product Name |
Edan IM50 patient monitor parameter panel board
|
Brand | Edan |
Use for | Edan IM50 patient monitor |
Material | Plastic |
Condition | Used |
Warranty | 90 days |
MOQ | One piece |
Parameter Monitoring Failures
Phenomenon | Possible Cause | Solution |
No ECG waveform | Bad connection of ECG cable. | Check the connection of the ECG cable. |
The ECG cable is damaged | Replace the ECG cable | |
Bad connection of the electrodes. | Check the connection | |
Circuit failure of ECG on the main control board. | Replace the main control board | |
No SpO2 trace or measurements. | SpO2 sensor failure | Replace the SpO2 sensor |
Bad connection of the cable of the SpO2 sensor | Check the connection | |
Circuit failure of SpO2 on the main control board. | Replace the main control board | |
The cuff fails to be inflated | Bad connection of the cuff, the pump and the NIBP connector | Check the connection outside the monitor |
Bad connection of BP module and NIBP connector | Check the connection inside the monitor. | |
The cuff or the extended cable is damaged | Replace the damaged parts | |
Circuit failure of BP on the main control board | Replace the main control board | |
NIBP cuff pressure value has no obvious change. | Circuit failure of BP on the main control board. | Replace the main control board. |
NIBP tube leakage | Please perform leakage test and change the tube or valve. | |
No NIBP measurement values or no alarm information | Cuff or extension tube abnormal | Please replace the cuff or extension tube. |
Circuit failure of BP on the main control board. | Replace the main control board | |
No TEMP measurement value. | TEMP sensor failure. | Replace the TEMP sensor |
Bad connection of TEMP sensors. | Check the connection | |
Circuit failure of TEMP on the main control board | Replace the main control board. |