Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Parts

240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly

240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly

  • 240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly
  • 240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly
  • 240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly
  • 240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly
240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly
Product Details:
Brand Name: philip
Model Number: FM20 FM30
Payment & Shipping Terms:
Minimum Order Quantity: 1
Delivery Time: 3-5 working days
Payment Terms: T/T, Western Union,PayPal
Contact Now
Detailed Product Description
Stock: Available Now Product Name: Philip Recorder
Type: FM20 FM30 Application: Medical Equipment,ICU,hospital
Warranty: 90days Key Words: Recoder Board, Repair Parts
Functionality: ECG Machine Repairs Superiority: Specialized Engineers With 12 Years Of Experience
State: In Good Condition Range Of Nominal Tension: 110V—240v
Context Of Use: Hospital; Clinic Sale Protection: In The Warranty Period And Non-human Damage Can Go To The Sale Free Of Charge For New
Feature: High Performance Aim Of Management: Original Assembled Real Product, Quality Guarantee, Deliver On Time, Preference Price
High Light:

240v Hospital Fetal Monitor

,

philip Avalon FM20 Hospital Fetal Monitor

Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly

philip Avalon FM20 FM30 Fetal Monitor Printer Assembly

 

 

►Brief Introduction

The fetal recorder is located in the Top Cover Assembly. Checking the Fetal Recorder Offset: The easiest way to check the recorder offset is to connect a resting Toco transducer (one that is not under any load) to the monitor and then change the offset setting until the trace is recording 20 units on the paper. Due to the delay between changing the offset setting and seeing the change on the paper, you may have to repeat this procedure to set the offset.

When viewed from the front of the monitor, 0 is the setting that prints the trace the furthest to the right, and 10 is the setting that prints the trace furthest to the left. If the trace from the resting Toco transducer is printed below the 20 unit gridline, you need to increase the offset setting. If the trace is printed above the 20 unit gridline, you need to decrease the offset setting. When the trace is recording 20 units, the offset is correctly set.

 

 

Specification

Resolution 1/4 bpm
Paper Advance Up to 20 mm/s
Accuracy ± 5 mm/page

 

 

Fetal Recorder Maintenance

The recorder platen, thermal printhead and paper sensor should be cleaned at least once a year, or when needed (when traces become faint).

Clean the assemblies as follows:

l Clean the recorder platen with a lint-free cloth using a soap/water solution.

l Wipe the printhead using a cotton swab moistened with 70% Isopropyl alcohol based solution.

l Check the paper sensing mechanism is dust free.

 

 

Removing the Recorder Chassis

1 Remove the top cover assembly.

2 Slide open the paper drawer, to gain access to the two countersunk screws (in the following photographs, we have removed the drawer assembly).

3 Remove the two countersunk screws.

4 Turn over the top cover assembly and place it top down on a cloth or other soft surface.

5 First release the cable lock by moving it in the direction of the arrows, then disconnect the white recorder ribbon cable from the recorder adapter board.

240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly 0

6 Next, disconnect the silver display ribbon cable from the recorder adapter board after releasing the cable lock.

240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly 1

7 Disconnect the stepper motor the paper sensor cable connectors from the recorder adapter board.

240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly 2

8 Remove the two screws (one on each side) fastening the RFI bracket to the recorder chassis.

240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly 3

9 Partially unscrew the screw nearest the recorder adapter board holding the stepper motor to the recorder chassis, until the thread no longer protrudes. This allows the necessary clearance for removing the RFI bracket.

240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly 4

10 Remove the four screws holding the recorder chassis.

240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly 5240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly 6

11 Free the RFI bracket on the stepper motor side by applying a little sideways pressure to the recorder chassis, while at the same time pulling the RFI bracket forwards.

240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly 7

12 Remove the RFI bracket.

240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly 8

13 Lift out the recorder chassis.

240v Hospital Fetal Monitor philip Avalon FM20 FM30 Fetal Monitor Printer Assembly 9

 

 

 

 

 

Thank you for your time! "Quality assurance" is our eternal pursuit and promise.

Don't be hesitant to contact us for philip Avalon FM20 FM30 Fetal Monitor Printer Assembly. Welcome to place your orders.

 

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Rita

Tel: 8618279091952

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.