China World Technology Medical Equipment Service Group
Win-win together!
Brand Name: | Mindray |
Model Number: | 040APS15 B |
Minimum Order Quantity: | 1pcs |
---|---|
Packaging Details: | Carton |
Delivery Time: | 3-5 working days |
Payment Terms: | T/T(wire transfer), Western Union, Paypal |
Product Name: | 040APS15 B Mindary VS600 Vital Signs Patient Monitor Power Supply Board | Brand: | Mindray |
---|---|---|---|
Part No: | 040APS15 B | Use For: | Mindary VS600 Vital Signs Patient Monitor |
AC Line Voltage: | 100 To 240 VAC (±10 %) | AC Frequency: | 50/60 Hz (±3 Hz) |
AC Current: | 0.9 To 0.5 A | MOQ: | One Piece |
Warranty: | 90 Days | Delivery Method: | DHL,UPS ,Fedex |
Highlight: | VS600 Vital Signs Patient Monitor Parts,Power Supply Board Patient Monitor Parts,040APS15 B Power Supply Board |
040APS15 B Mindary VS600 Vital Signs Patient Monitor Power Supply Board
Specification
AC line voltage
|
100 to 240 VAC (±10 %)
|
AC frequency
|
50/60 Hz (±3 Hz)
|
AC current
|
0.9 to 0.5 A
|
Power management and interface board
The power management interface card mainly performs the following functions:
Power On/Off Failure
Symptoms
|
Possible Cause
|
Troubleshooting
|
The monitor fails to start.
|
AC mains not connected or battery too low
|
Check that AC mains is properly connected or battery capacity is sufficient.
|
Cable defective
|
1. Check if the cable between the keyboard and motherboard is connected properly. 2. Verify that the cables between the power board and the power management board are properly connected. 3. Verify that the cables between the motherboard and the power management board are properly connected. |
|
Power board defective
|
Replace the power board.
|
|
Power management board defective
|
Replace the power management board.
|
|
The main board failed.
|
Replace the main board
|
Battery Failures
Symptoms
|
Possible Cause
|
Troubleshooting
|
Battery cannot be charged
|
Battery defective
|
Replace the battery.
|
Cable defective
|
Check that the cable between the battery interface board and power management board is correctly connected.
|
|
Power management board defective
|
Replace the power management board.
|
|
Battery interface board defective
|
Replace the battery interface board.
|
Disassembling AC/DC Power Board and Battery Converter Board
1.Unscrew the four M3×6 screws as indicated and take out the AC/DC power board
2.Unscrew the two M3 nuts to take out the battery interface board.
Our philosophy
self-cultivation, family harmony, career development, and helping the world.
Customer service
Our goal is to provide excellent customer service so that your relationship with us meets or exceeds your expectations every time.
All members of our company strive to answer our customers' questions or concerns.
Tell us what you're looking for and we'll let you know when we get the medical supplies you need.
Whether you have questions, comments or compliments, we'd love to hear them.
Thank you so much for your attention .If you're interested in 040APS15 B Mindary VS600 Vital Signs Patient Monitor Power Supply Board ,feel free to contact us .Thank you .
Contact Person: Kiara
Tel: 8619854815217