Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Parts

PN 115-028051-00 Patient Monitor Parts Mindray SV300 600 650 800 850 Ventilator Blower Turbine Box Assembly

PN 115-028051-00 Patient Monitor Parts Mindray SV300 600 650 800 850 Ventilator Blower Turbine Box Assembly

  • PN 115-028051-00 Patient Monitor Parts Mindray SV300 600 650 800 850 Ventilator Blower Turbine Box Assembly
  • PN 115-028051-00 Patient Monitor Parts Mindray SV300 600 650 800 850 Ventilator Blower Turbine Box Assembly
  • PN 115-028051-00 Patient Monitor Parts Mindray SV300 600 650 800 850 Ventilator Blower Turbine Box Assembly
  • PN 115-028051-00 Patient Monitor Parts Mindray SV300 600 650 800 850 Ventilator Blower Turbine Box Assembly
PN 115-028051-00 Patient Monitor Parts Mindray SV300 600 650 800 850 Ventilator Blower Turbine Box Assembly
Product Details:
Brand Name: Mindray
Model Number: SV300 600 650 800 850
Payment & Shipping Terms:
Minimum Order Quantity: 1
Delivery Time: 3-5 working days
Payment Terms: T/T, Western Union,PayPal
Contact Now
Detailed Product Description
Product Name: Mindray Type: SV300 600 650 800 850
Warranty: 90days Applications Scenes: Ventilator
Product Status: In Good Working Condition. Stock Status: Stock Available
Minimum Purchasing Quantity: 1unit O2 %: ± (3 Vol.% +1 % Of Setting)
TV: ± (10 ML + 10 % Of Setting) (BT Tinsp: ± 0.1 S Or ± 10 % Of Setting, Whichever Is Greater
I: E: 2: 1 To 1: 4: ± 10 % Of Setting, Other Range: ± 15% Of Setting F: ± 1 Bpm
High Light:

PN 115-028051-00 Patient Monitor Parts

,

Mindray SV300 Patient Monitor Parts

Mindray SV300 600 650 800 850 Ventilator Machine Blower box Turbine Box assembly PN 115-028051-00 

 

 

 

► Brief Introduction

The pneumatic circuit system includes four parts, namely the inspiratory limb, patient tubing, expiratory limb and backup air supply module. The inspiratory limb system can be disassembled into the gas supply subsystem, flow control subsystem, safety valve subsystem and nebulization control subsystem.

PN 115-028051-00 Patient Monitor Parts Mindray SV300 600 650 800 850 Ventilator Blower Turbine Box Assembly 0

As shown above, patient tubing is the bridge connecting the inspiratory limb and the expiratory limb. Meanwhile, the gas supply and the nebulization subsystems also can be connected to the patient tubing to perform the nebulization function. These systems together constitute a closed circuit to conduct ventilation management functions.

The gas supply subsystem is the first part of the pneumatic circuit, and its main function is to draw the external air and oxygen supplys into the machine. As the pressure of the external gas supply is relatively high and unstable, with certain impurities, a 5 um filter has been provided in the air supply subsystem. As the pressure of the external oxygen supply is relatively high and unstable, a pressure regulation valve has been provided in the air supply subsystem to protect the precision solenoid valve and flow sensor in the flow control module at the back-end. Meanwhile, a pressure sensor has been provided to monitor the air supply pressure, as well as a check valve to prevent the reflux of the gas inside the ventilator to other external air supply.

The gas supply inlet connector is designed as NIST/DISS optional type to prevent incorrect connection. The pressure regulation valve at the back-end of the gas supply inlet reduces the gas supply pressure to ensure the output of the proportional solenoid valve PSOL is stable and maintains relatively good reproducibility.

 

 

 

 

 

 

 

Control Accurac

O2 % ± (3 vol.% +1 % of setting)
TV ± (10 mL + 10 % of setting) (BT
Tinsp ± 0.1 s or ± 10 % of setting, whichever is greater
I: E 2: 1 to 1: 4: ± 10 % of setting, other range: ± 15% of setting
f ± 1 bpm
fSIMV ± 1 bpm
Tslope ± (0.2 s + 20 % of setting)
PEEP ± (2.0 + 5 % of setting) cmH2O
∆Pinsp ± (2.0 + 5 % of setting) cmH2O
∆Psupp ± (2.0 + 5 % of setting) cmH2O
Phigh ± (2.0 + 5 % of setting) cmH2O
Plow ± (2.0 + 5 % of setting) cmH2O
Thigh ± 0.2 s or ± 10 % of setting, whichever is greater
Tlow ± 0.2 s or ± 10 % of setting, whichever is greater
Pressure Trigger ± (1.0 + 10 % of setting) cmH2O
Flow Trigger ± (1.0 L/min + 10 % of setting)
int.PEEP ∆ ± (2.0 + 5% of setting) cmH2O
Exp % ± 10 %
fapnea ± 1 bpm
Papnea ∆ ± (2.0 + 5 % of setting) cmH2O
TVapnea ± (10 mL + 10 % of setting) (BTPS)
Apnea Tinsp ± 0.1 s or ± 10% of setting, whichever is greater

 

 

 

 

Please kindly contact us, if you are interested in Mindray SV300 600 650 800 850 Ventilator Machine Blower box Turbine Box assembly PN 115-028051-00 , Thank you!

 

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Rita

Tel: 8618279091952

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.