Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Parts

Welch Allyn Vital Sign Monitor 300 Series 53NTP Recorder Printer REV A 630-0213-10

Welch Allyn Vital Sign Monitor 300 Series 53NTP Recorder Printer REV A 630-0213-10

  • Welch Allyn Vital Sign Monitor 300 Series 53NTP Recorder Printer REV A 630-0213-10
  • Welch Allyn Vital Sign Monitor 300 Series 53NTP Recorder Printer REV A 630-0213-10
  • Welch Allyn Vital Sign Monitor 300 Series 53NTP Recorder Printer REV A 630-0213-10
  • Welch Allyn Vital Sign Monitor 300 Series 53NTP Recorder Printer REV A 630-0213-10
Welch Allyn Vital Sign Monitor 300 Series 53NTP Recorder Printer REV A 630-0213-10
Product Details:
Brand Name: Welch Allyn
Model Number: 300 Series 53NTP
Payment & Shipping Terms:
Minimum Order Quantity: 1
Delivery Time: 3-5 working days
Payment Terms: T/T, Western Union,PayPal
Contact Now
Detailed Product Description
Product Name: Welch Allyn Vital Signs Monitor 300 Series 53NTP PN: 300 Series 53NTP
Application: Medical Equipment,ICU,hospital Device Warranty: 90days
Color: White Operating Temperature: +10 °C To +40 °C +50 °F To +104 °F
Operating Altitude: -170 M To +4877 M -557 Ft To + 16,000 Ft Shipping Altitude: -615 M To 12,300 M -2000 Ft To 40,000 Ft
Storage Temperature: -20 °C To +50 °C -4 °F To +122 °F Relative Humidity: 15% To 90% (non-condensing)
Printer Current: Peak (mA): 0.940 MA
High Light:

630-0213-10 Vital Sign Monitor

,

53NTP Recorder Printer Vital Sign Monitor

Welch Allyn Vital Signs Monitor 300 Series 53NTP Recorder Printer REV A 630-0213-10

In Good working Condition Welch Allyn Vital Signs Monitor 300 Series 53NTP Recorder Printer REV A 630-0213-10

 

► Brief Introduction

The printer provides a way to view and save patient vital-signs data. It can be configured to print all stored data as a single batch or to print a continuing stream of data as the data is recorded in memory.

The printer standard report consists of a header, patient information, patient data (table heading, date, time, range and alarm flags, and measurement readings), an error legend and a footer. If the monitor receives a print request and no data is available, it prints only the header and footer.

 

 

 

► To Start and Stop Printing

• If the monitor is not printing, press button to start printing.

• If the monitor is printing, press button to stop printing. (If the monitor is set to Stream

printing, the printer prints a footer before stopping.)

 

 

 

► Functional Verification

1. Put the monitor into Service Mode.

2. Verify that the printer has paper.

3. Verify that a settings report prints, and that it contains no printed anomalies and no missing or faded sections.

4. After 2 seconds of printing, verify that the current draw is no more than 2.5A above the baseline current recorded in step 2 of the Baseline Current Draw verification test.

5. Exit the Service Mode

 

With a new roll of paper, the first line might be faded. This does not indicate a problem.

The settings report contains a calibration record, user record, hardware status record, and software versions record.

• The calibration record includes manufacturing configuration data: monitor serial number, set parameters, and language.

• The user record includes user-configurable settings: alarm limits, patient type, parameter modes/units, auto interval, tone volume, and others.

• The hardware status record shows the hardware revision number, the battery voltage level, the total number of NIBP monitoring cycles completed on the monitor, and the total number of hours of operation of the monitor.

• The software versions record indicates the software version numbers of the main board, SpO 2 and Temperature options (if present), and NIBP.

 

 

 

► Quick Details

Brand Welch Allyn
Type 300 Series 53NTP
Warranty 90 Days
Shipping Date 3-5 Working Days
Application Hospital Devices; Medical Care; ICU; Medical Equipments and Accessories

 

 

 

► Remove and Disassemble the Printer

1. Remove the printer door.

Open the printer door and remove the roll of printer paper.

Welch Allyn Vital Sign Monitor 300 Series 53NTP Recorder Printer REV A 630-0213-10 0

Push the printer door toward the rear of the monitor until it snaps out of the two support spindles on the printer frame.

Welch Allyn Vital Sign Monitor 300 Series 53NTP Recorder Printer REV A 630-0213-10 1

2. Remove the printer assembly.

Push the plastic legs of the printer assembly outward and pull the printer assembly forward to free it from the snap-in seating of the rear chassis bulkhead.

 

 

 

 

Thank you for your time! If you are interested in Welch Allyn Vital Signs Monitor 300 Series 53NTP Recorder Printer REV A 630-0213-10

, please let us know. Welcome to place your orders.

Thank you for your time and interest.

 

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Rita

Tel: 8618279091952

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.