China World Technology Medical Equipment Service Group
Win-win together!
Brand Name: | Covidien |
Model Number: | Puritan Bennett 840 |
Minimum Order Quantity: | 1 |
---|---|
Packaging Details: | ► Turn the wheel to highlight the number of records to print. The choices are: |
Delivery Time: | 3-5 working days |
Payment Terms: | T/T, Western Union,PayPal |
Product Name: | Covidien | Type: | Puritan Bennett 840 |
---|---|---|---|
Application: | Medical Equipment,ICU,hospital Device | Warranty: | 90days |
Temperature Range Operating: | 10 To 40 °C (50 To 104 °F) At 10 To 95% Relative Humidity, Noncondensing | Temperature Range Storage: | -20 To 50 °C (-4 To 122 °F) At 10 To 95% Relative Humidity, Noncondensing |
Atmospheric Pressure Operating: | 700 To 1060 HPa (10.2 To 15.4 Psi) | Atmospheric Pressure Storage: | 500 To 1060 HPa (7.3 To 15.4 Psi) |
Altitude Range Operating: | -443 To 3280 M (-1350 To 10,000 Ft) | Altitude Range Storage: | Up To 6560 M (20,000 Ft) |
Highlight: | PN 4-072212-SP,Medical Devices Exhalation Flow Sensor,Medical Equipment Exhalation Flow Sensor |
Covidien Puritan Bennett 840 ventilator Exhalation Flow Sensor Q3 PN 4-072212-SP
► Brief Introduction
The exhalation flow sensor (Q3) provides flow information on exhaled gas. Flow sensor measurements are used to determine net gas flow to the patient and spirometry
During exhalation, patient gas enters the exhalation system from the patient circuit. The ventilator actively controls the exhalation valve (EV), opening and closing it in precise increments throughout inspiration and exhalation. Such precise control of the valve allows the ventilator to deliver aggressive breaths while minimizing pressure overshoots, controlling PEEP, and relieving excess pressures. The exhalation module also performs spirometry of the exhaled gas and monitors flow sensitivity.
The exhalation heater (EXH HTR) heats the exhaled gas to minimize moisture in the exhalation module, while thermistors provide the ventilator with temperature feedback. The expiratory pressure transducer (PE), on the exhalation transducer PCB, and the exhalation flow sensor (Q3) provide readings used in breath delivery alculations.
► Installing exhalation flow sensor (Q3)
1 If you are installing a new Q3, remove caps from both ends.
2 Make sure O-rings at both ends of Q3 tube are in place.
3 Orient body of Q3 with harness connection facing up.
4 Gently insert longer portion of tube into heater port.
5 Connect cable to connector on Q3.
6 Install exhalation valve.
► Running Flow Sensor Calibration
1 Enter Service Mode.
2 Ensure both air and oxygen are connected to the ventilator.
3 Connect the gold standard circuit between the To Patient and From Patient ports.
4 Ensure that there are no leaks or occlusions by running EST and ensure that the ventilator passes the EST Leak Test. Once this test has passed, exit EST.
5 Select the Other Screens button.
6 Select Flow Sensor Calibration.
7 Press the Start button and press Accept.
The ventilator automatically starts the calibration. When the calibration is complete, press the Other Screens button again to return to the Other Screens menu.
► Specifications
Temperature range Operating | 10 to 40 °C (50 to 104 °F) at 10 to 95% relative humidity, noncondensing |
Temperature range Storage | -20 to 50 °C (-4 to 122 °F) at 10 to 95% relative humidity, noncondensing |
Atmospheric pressure Operating | 700 to 1060 hPa (10.2 to 15.4 psi) |
Atmospheric pressure Storage | 500 to 1060 hPa (7.3 to 15.4 psi) |
Altitude range Operating | -443 to 3280 m (-1350 to 10,000 ft) |
Altitude range Storage | Up to 6560 m (20,000 ft) |
► Product Application
We specialize in the sales and maintenance of various medical equipment and accessories. Our main brands are philip, Mindray, GE, Goldway, NIHON KOHDEN, etc. Since the establishment of the company, we have explored and developed in this industry for many years. Our products are mainly used for medical equipment in hospitals or clinics. We sell complete machines, spare parts, and also develop product maintenance services. Our main products are patient monitors, defibrillators, ECG machines and various module accessories. If you want to know more detailed information about our products, welcome to contact us. We will give you a detailed introduction.
Thank you for your time! If you are interested in Covidien Puritan Bennett 840 ventilator Exhalation Flow Sensor Q3 PN 4-072212-SP, please let us know. Welcome to place your orders.
Thank you for your time and interest.
Contact Person: Bella Liu
Tel: +8618879075074