Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsUsed Patient Monitor

Mindray IMEC8 Patient Monitor Display 8.4'' TM084SDHG01 M1P6563706400

Mindray IMEC8 Patient Monitor Display 8.4'' TM084SDHG01 M1P6563706400

  • Mindray IMEC8 Patient Monitor Display 8.4''  TM084SDHG01 M1P6563706400
  • Mindray IMEC8 Patient Monitor Display 8.4''  TM084SDHG01 M1P6563706400
  • Mindray IMEC8 Patient Monitor Display 8.4''  TM084SDHG01 M1P6563706400
  • Mindray IMEC8 Patient Monitor Display 8.4''  TM084SDHG01 M1P6563706400
Mindray IMEC8 Patient Monitor Display 8.4''  TM084SDHG01 M1P6563706400
Product Details:
Brand Name: Mindray
Model Number: TM084SDHG01 M1P6563706400
Payment & Shipping Terms:
Minimum Order Quantity: 1pcs
Packaging Details: Carton
Contact Now
Detailed Product Description
Delivery Time: 3-5 Working Days Payment Terms: Full Before The Shipment
Product Name: Mindray IMEC8 Patient Monitor Display 8.4" TM084SDHG01 M1P6563706400 Brand: Mindray
Part No: TM084SDHG01 M1P6563706400 MOQ: One Piece
Warranty: 90 Days Condition: Used
Suitable For: IMEC8 Delivery Method: DHL ,UPS ,Fedex ,etc
Payment Method: T/T ,Paypal ,Western Union Size: 8.4"
Resolution: 800 X 600 Pixels Waveform: Up To 8
High Light:

Mindray IMEC8 Patient Monitor

,

M1P6563706400 Patient Monitor

,

TM084SDHG01 Patient Monitor

Mindray IMEC8 Patient Monitor Display 8.4" TM084SDHG01 M1P6563706400

 

 

The iMEC8, iMEC10 and iMEC12 support 8", 10" and 12" LCD monitors respectively. The displays are connected to the motherboard via display cable and backlight cable. The connection cables of the three patient monitors are different and unreliable. The signal cable is pretty bad. If there is a problem with the display Consider replacing the display cable first.

 

Touch screen control panel and touch screen

Fully configurable iMEC patient monitor supports touch screen. The touch screen can be controlled by touch and can be calibrated. The touch screen is connected to the touch screen control panel and the main board. The touch screen control panel operates on the touch screen and allows communication with the patient monitor. The touch screen control panel communicates with the main board through the I2C interface, and the touch screen is connected via FPC.

 

Specification

 

Size 8.4”
Resolution 800 x 600 pixels
Dimensions 268 x 210 x 114mm
Waveform Up to 8
External display 1 display through VGA

 

 

 

Display Failures

 

Symptoms Possible Cause Corrective Action

 

 

The display is

blank or black.

 

 

Cable defective

1. Check whether the cable between the monitor and the motherboard and the backlight cable are connected correctly.

2. Inspect cables and connectors for damage.

Main board defective Replace the main board
isplay defective Replace the display.

 

Symptoms Possible Cause Corrective Action

Images

overlapped or

distorted

 

main board error

Replace the main board, or upgrade the main

board with the upgrade software.

 

Cable defective

1. Check whether the cable between the monitor and the motherboard and the backlight cable are connected correctly.

2. Inspect cables and connectors for damage.

 

Symptoms Possible Cause Corrective Action

Snows or flashing

specks on the

display.The display does

not function.

 

 

Cable defective

1. Check that the cable between the secondary monitor and the monitor is properly connected.

 

2. Inspect cables and connectors for damage.

The connector board

failed.

 

Replace the connector board .
The main board failed. Replace the main board.

 

Mindray IMEC8 Patient Monitor Display 8.4''  TM084SDHG01 M1P6563706400 0

 

Symptoms Possible Cause Corrective Action

Touchscreen does

not respond

 

Touchscreen disabled

Check if there is a Mindray IMEC8 Patient Monitor Display 8.4''  TM084SDHG01 M1P6563706400 1

displayed above

the [

Main Menu

] quickkey. If yes, press and hold

the [

Main Menu

] quickkey for more than 3

seconds to enable the touchscreen

 

Cable defective

1.Check whether the cables between the touch screen and the touch screen control board, and between the touch screen control board and the main board are connected correctly.

2. Check for loose cables and connectors.

Touchscreen control

board defective

 

Replace the touchscreen control board
Touchscreen defective Replace the touchscreen.
The main board failed. eplace the main board.

Touch position

invalid

 

Touchscreen not

calibrated

 

Calibrate the touchscreen.

 

 

If you're interested in Mindray IMEC8 Patient Monitor Display 8.4" TM084SDHG01 M1P6563706400 ,feel free to contact us .Thank you .

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Kiara

Tel: 8619854815217

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.