China World Technology Medical Equipment Service Group
Win-win together!
Brand Name: | Philip |
Model Number: | MP70 |
Minimum Order Quantity: | 1 |
---|---|
Delivery Time: | 3-5 working days |
Payment Terms: | T/T, Western Union,PayPal |
Product Name: | Philip MP70 ECG Out Board | Type: | M8085-67001 |
---|---|---|---|
Application: | Patient Monitor | Warranty: | 90days |
Color: | Green | Product Edge: | Original |
Major Materials: | PCB,Plastic,etc | Product Labels: | Patient Monitor Network Card |
Apply For: | Patient Monitor | Scenarios: | ICU, Emergency, Patient Monitor, ICU Room And Operating Room,etc |
Product Situation: | Running Sweetly | Range Of Nominal Tension: | 110V—240v |
Features Introduction: | Durable In Use,Professional, Reliable | Stock Situation: | Sufficient Inventory |
Minimum Purchasing Quantity: | 1pcs | ||
High Light: | M8085-67001 Patient Monitor Parts,ECG Out Board Patient Monitor Parts,M8085-67001 Patient Monitor CPU Board |
Hospital Medical Equipment Parts Philipilip MP70 Patient Monitor ECG Out Board M8085-67001 CPU Board
► Brief Introduction
ECG-Out Marker-In
ECG Out/Marker In interface Receives ECG waveforms directly from the ECG/Resp arrhythmia ST-Segment Philipilipysiological algorithm through the RS-422 serial interface and converts digital ECG signals to analog ECG signals. In addition, the ECG Out controller receives labeling information from the connected device and forwards this data to the ECG/Resp arrhythmia ST segment Philipilipysiological algorithm. The converted analog signal is used to synchronize the connected device with the patient's electrocardiogram
RS-422
The SERIAL link RS-422 interface transmits the ECG signal to the ECG output/label input of the IntelliVue patient monitoring system display unit. The interface is a serial, differential, full-duplex link. The interface has ESD protection.
PS/2
The PS/2 interface supports the serial protocol for standard PS/2 devices (mice). PS/2 serial protocol is interpreted by the HIF control interface.
► Removing the ECG Out board
For instructions on removing the ECG Out board when an SRL2 board is plugged
1 Remove the mounting plate from the monitor if attached.
2 Remove the four screws with a T20 screwdriver and pull off the bottom cover.
3 Unscrew the screws on the ECG Out board and pull on the tab to remove the cable.
► Troubleshooting the ECG OUT
Cause of Failure | Failure Isolation and Remedy |
No marker pulse is displayed on the Monitor or no ECG- OUT signal to the Defib |
Disconnect the MMS and Defib cable. Switch the Monitor off then on again. Observe the red LED in the ECG OUT section. (Note that the LED can only be observed if the housing bottom is removed). |
ECG OUT Board defective | Replace ECG OUT board |
ECG OUT Board defective | If the red LED switches on and remains on for more than 20 seconds: Replace ECG OUT Board |
ECG OUT Board defective | Connect Known good Defib, Defib cable MMS and MMS cable. Check Marker pulse and ECG OUT signal at defib again. If there is still no signal: Replace ECG OUT board |
► ECG Out Functionality
Connections
The cable M1181A #A62 has both ends terminated. The PhilipilipotograPhilipilip above shows the monitor side
connection.
If using a non-terminated cable:
► Delivery Clauses
To make sure the work goes smoothly, please pay special attention to him and arrange your time to apply carefully according to your personal needs.For delivery time, we will do our best to process and ship orders as soon as possible. Please be patient and stay in touch with us. In general, most orders will be shipped within 3 to 5 business days and can arrive on time. However, during busy days, the shipment may be delayed by more than 5 business days. If you are in dire need of your product, please be sure to confirm with us before submitting your applications. If your order takes longer than expected or if you want to clarify that upon arrival, feel free to contact us. We're always here to help you.