Hospital Patient Monitor Parts ECG Out Board M8085-67001 CPU Board

Hospital Patient Monitor Parts ECG Out Board M8085-67001 CPU Board

  • Hospital Patient Monitor Parts ECG Out Board M8085-67001 CPU Board
  • Hospital Patient Monitor Parts ECG Out Board M8085-67001 CPU Board
  • Hospital Patient Monitor Parts ECG Out Board M8085-67001 CPU Board
  • Hospital Patient Monitor Parts ECG Out Board M8085-67001 CPU Board
Hospital Patient Monitor Parts ECG Out Board M8085-67001 CPU Board
Product Details:
Brand Name: Philip
Model Number: MP70
Payment & Shipping Terms:
Minimum Order Quantity: 1
Delivery Time: 3-5 working days
Payment Terms: T/T, Western Union,PayPal
Contact Now
Detailed Product Description
Product Name: Philip MP70 ECG Out Board Type: M8085-67001
Application: Patient Monitor Warranty: 90days
Color: Green Product Edge: Original
Major Materials: PCB,Plastic,etc Product Labels: Patient Monitor Network Card
Apply For: Patient Monitor Scenarios: ICU, Emergency, Patient Monitor, ICU Room And Operating Room,etc
Product Situation: Running Sweetly Range Of Nominal Tension: 110V—240v
Features Introduction: Durable In Use,Professional, Reliable Stock Situation: Sufficient Inventory
Minimum Purchasing Quantity: 1pcs
High Light:

M8085-67001 Patient Monitor Parts

,

ECG Out Board Patient Monitor Parts

,

M8085-67001 Patient Monitor CPU Board

 

Hospital Medical Equipment Parts Philipilip MP70 Patient Monitor ECG Out Board M8085-67001 CPU Board

 

 

 

► Brief Introduction

ECG-Out Marker-In

ECG Out/Marker In interface Receives ECG waveforms directly from the ECG/Resp arrhythmia ST-Segment Philipilipysiological algorithm through the RS-422 serial interface and converts digital ECG signals to analog ECG signals. In addition, the ECG Out controller receives labeling information from the connected device and forwards this data to the ECG/Resp arrhythmia ST segment Philipilipysiological algorithm. The converted analog signal is used to synchronize the connected device with the patient's electrocardiogram

RS-422

The SERIAL link RS-422 interface transmits the ECG signal to the ECG output/label input of the IntelliVue patient monitoring system display unit. The interface is a serial, differential, full-duplex link. The interface has ESD protection.

PS/2

The PS/2 interface supports the serial protocol for standard PS/2 devices (mice). PS/2 serial protocol is interpreted by the HIF control interface.

 

 

Removing the ECG Out board

For instructions on removing the ECG Out board when an SRL2 board is plugged

1 Remove the mounting plate from the monitor if attached.

2 Remove the four screws with a T20 screwdriver and pull off the bottom cover.

Hospital Patient Monitor Parts ECG Out Board M8085-67001 CPU Board 0

3 Unscrew the screws on the ECG Out board and pull on the tab to remove the cable.

Hospital Patient Monitor Parts ECG Out Board M8085-67001 CPU Board 1

 

 

Troubleshooting the ECG OUT

Cause of Failure Failure Isolation and Remedy
No marker pulse is
displayed on the Monitor
or no ECG- OUT signal to
the Defib
Disconnect the MMS and Defib cable.
Switch the Monitor off then on again.
Observe the red LED in the ECG OUT section. (Note that the LED can only be observed if the housing bottom is removed).
ECG OUT Board defective Replace ECG OUT board
ECG OUT Board defective If the red LED switches on and remains on for more than 20 seconds:
Replace ECG OUT Board
ECG OUT Board defective Connect Known good Defib, Defib cable MMS and MMS cable. Check Marker pulse and ECG OUT signal at defib again. If there is still no signal:
Replace ECG OUT board

 

 

ECG Out Functionality

Connections

Hospital Patient Monitor Parts ECG Out Board M8085-67001 CPU Board 2

The cable M1181A #A62 has both ends terminated. The PhilipilipotograPhilipilip above shows the monitor side

connection.

If using a non-terminated cable:

  1. Strip 5 mm (3/16") insulation from leads and twist conductor strands tightly.
  2. Solder leads to the connector as shown in the following diagram.

 

 

► Delivery Clauses

To make sure the work goes smoothly, please pay special attention to him and arrange your time to apply carefully according to your personal needs.For delivery time, we will do our best to process and ship orders as soon as possible. Please be patient and stay in touch with us. In general, most orders will be shipped within 3 to 5 business days and can arrive on time. However, during busy days, the shipment may be delayed by more than 5 business days. If you are in dire need of your product, please be sure to confirm with us before submitting your applications. If your order takes longer than expected or if you want to clarify that upon arrival, feel free to contact us. We're always here to help you.

 

 

Thank you for your time and interest. If you have any questions about Philipilip MP70 Patient Monitor ECG Out Board M8085-67001, please feel free to contact us at any time, we are glad to answer them. Customer satisfaction. Our biggest pursuit!

 

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Rita

Tel: 8618279091952

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.