China World Technology Medical Equipment Service Group
Win-win together!
| Place of Origin: | Japan |
| Brand Name: | Nihon Kohden |
| Certification: | N/A |
| Model Number: | LQ121S1LG46 |
| Minimum Order Quantity: | 1 Piece |
|---|---|
| Price: | USD |
| Packaging Details: | Standard Export Packing |
| Delivery Time: | 3-5 working days |
| Payment Terms: | T/T,Western Union,Paypal |
| Supply Ability: | 10pcs |
| Brand: | Nihon Kohden | Product Name: | Display |
|---|---|---|---|
| Applicable Model: | BSM-4113K | Model: | LQ121S1LG46 |
| Display Type: | TFT LCD Display | Size: | 12.1 Inch |
| Viewing Area: | 246 Mm X184.5mm | Resolution: | 800x600 Dots |
| Waveform Display Mode: | Non-fade Moving Or Non-fade Fixed | Sweep Time: | 7.4 S (at 25 Mm/s Sweep Speed) |
| Waveform Display Color: | 12 | Numeric Display Color: | 12 |
| Transportation: | DHL, FedEx, UPS, TNT, EMS, By Air, Etc. | MOQ: | One Piece |
| Highlight: | 12.1 inch TFT LCD display,LQ121S1LG46 LCD monitor replacement,Nihon Kohden BSM-4113K display part |
||
BSM-4113K Monitor Display
12.1 Inch TFT LCD Display LQ121S1LG46 for Nihon Kohden BSM-4113K Patient Monitor
► Brief Description
The BSM-4113K is equipped with a 12.1-inch high-resolution color LCD (TFT LCD). This screen is the device's most prominent hardware feature, designed to provide healthcare professionals with clear, real-time vital sign monitoring data. Its large size allows it to display more waveform channels and parameter values simultaneously, making it ideal for continuous monitoring when mounted on a bedside table or wall.
The design emphasizes ease of observation and convenient operation. The large 12.1-inch screen not only displays routine data but also supports trend reviews and alarm event lists, helping healthcare professionals quickly understand changes in a patient's condition over a period of time.
► Detailed Photos
![]()
![]()
► Features
► Troubleshooting
一,Fault Symptom: No display and the backlight does not light.
1. Possible Cause: Faulty connection between CN108 on the main board and inverter.
Solution: Check the connections.
2. Possible Cause: Faulty connection between inverter and LCD unit.
Solution: Check the connections.
3. Possible Cause: Faulty inverter board.
Solution: Replace the inverter board.
4. Possible Cause: Faulty backlight lamps.
Solution: Replace the backlight lamps.
二,Fault Symptom: Fine vertical or horizontal lines on the LCD.
1. Possible Cause: Faulty LCD unit.
Solution: Replace the LCD unit.
2. Possible Cause: Faulty MAIN board.
Solution: Replace the MAIN board.
二,Fault Symptom: The display is dim and the brightness control does not make it brighter.
1. Possible Cause: Battery operation.
Solution: Set the POWER SAVE MODE to OFF on the SYSTEM SETUP screen.
2. Possible Cause: Dirt on the touch screen (inside or outside).
Solution: Clean the surface of touch screen.
3. Possible Cause: The backlight lamps are near the end of its useful life. 25,000 hours for revision no.before CA 50,000 hours for revision no.CA or later
Solution: Replace the backlight lamps.
4. Possible Cause: Faulty connection between CN108 on the MAIN board and inverter board.
Solution: Check the connection.
Contact Person: Jasmine
Tel: 86-15180047496