logo
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Parts

Biocare IM12 Patient Monitor Parts Power Supply Board 02045052 E204460 94V-0

Biocare IM12 Patient Monitor Parts Power Supply Board 02045052 E204460 94V-0

  • Biocare IM12 Patient Monitor Parts Power Supply Board 02045052 E204460 94V-0
  • Biocare IM12 Patient Monitor Parts Power Supply Board 02045052 E204460 94V-0
  • Biocare IM12 Patient Monitor Parts Power Supply Board 02045052 E204460 94V-0
  • Biocare IM12 Patient Monitor Parts Power Supply Board 02045052 E204460 94V-0
  • Biocare IM12 Patient Monitor Parts Power Supply Board 02045052 E204460 94V-0
  • Biocare IM12 Patient Monitor Parts Power Supply Board 02045052 E204460 94V-0
Biocare IM12 Patient Monitor Parts Power Supply Board 02045052 E204460 94V-0
Product Details:
Brand Name: Biocare
Model Number: 02045052 E204460
Payment & Shipping Terms:
Minimum Order Quantity: 1
Delivery Time: 3-5 working days
Payment Terms: T/T, Western Union,Paypal
Contact Now
Detailed Product Description
Product Name: Power Supply Board Part No: 02045052 E204460
Machine Application: Biocare IM12 Patient Monitor (AC) Input Voltage: 100 V~240 V
Input Power: 75 VA Frequency: 50 Hz/60 Hz (Allowable Frequency Error ±1Hz)
Fuse: 3.15A/250V Safety Classification: Class I, Type BF, CF
Operating Temperature: 0°C - 40°C Warranty: 90 Days
Highlight:

Biocare IM12 power supply board

,

Patient monitor replacement power board

,

94V-0 rated monitor parts

Biocare IM12 Patient Monitor Parts Power Supply Board 02045052 E204460 94V-0

 

Brief Introduction:

 

This power board provides stable power to the Biocare IM12. Key components include: a yellow high-frequency transformer responsible for primary and secondary voltage conversion; multiple black electrolytic capacitors and copper-colored inductors forming a filtering circuit to ensure stable output voltage; and black heatsinks closely attached to heat-generating components, working in conjunction with the board's copper-clad heatsink design to prevent overheating. It converts external AC power or internal battery DC power into multiple precise low-voltage DC currents through transformers, capacitors, and other components to power the motherboard, display, and other components. It also integrates overvoltage and overcurrent protection and battery charge/discharge management functions.

 

Biocare IM12 Patient Monitor Parts Power Supply Board 02045052 E204460 94V-0 0

Power Supply Failure
 
Symptoms Possible Cause Solution

 

No +5V output/

No +12V output

 

1. Power supply protection

2.Power board failure

1. Turn off the monitor thenrestart it.
2. If the problem remains,disconnect the AC mains for 5 s and reconnect it, then restart the patient monitor.

3. If the problem still remains,replace power board.

No +3.3V output

1. Power board failure

2. Main board failure

1. Replace power board

2. Replace main board

 

AC Power Supply Connection:

 

Before connecting the AC power supply, first check the status of the monitor and its associated modules. Follow these steps to connect the AC power:
  • Ensure the AC power supply meets the following specifications: AC 100V~240V, 50 Hz/60 Hz. Use the power cord provided with the monitor—plug one end into the monitor’s power interface, and the other end into a single-phase outlet with protective grounding.
  • Use the dedicated grounding wire supplied with the monitor to connect the device to the protective ground terminals.
Special note: Ensure the monitor is properly grounded at all times.
 

Disassembling the Power Board

 
1. Remove the four M3*8 screws from the back cover by unscrewing them completely.
 
Biocare IM12 Patient Monitor Parts Power Supply Board 02045052 E204460 94V-0 1
 
2. Separate the front and back covers, then disconnect the connecting cables for the screen and button board.
 
Biocare IM12 Patient Monitor Parts Power Supply Board 02045052 E204460 94V-0 2
 
3.To remove the power board: Unscrew the three M3*6 screws and two ST2.9*8 screws completely. When lifting the power board out, be sure to disconnect all power connection cables first.
 
Biocare IM12 Patient Monitor Parts Power Supply Board 02045052 E204460 94V-0 3
 
 

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Rita

Tel: +8618279091952

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2026 China World Technology Medical Equipment Service Group. All Rights Reserved.