China World Technology Medical Equipment Service Group
Win-win together!
| Brand Name: | Biocare |
| Model Number: | LQ121S1LG55 |
| Minimum Order Quantity: | 1 Piece |
|---|---|
| Price: | USD |
| Packaging Details: | Standard Export Packing |
| Delivery Time: | 3-5 working days |
| Payment Terms: | T/T,Western Union,Paypal |
| Brand: | Biocare | Product Category: | Monitorning |
|---|---|---|---|
| Product Type: | Spare Parts | Type: | TFT LCD Screen |
| Item Name: | Display | Model: | LQ121S1LG55 |
| Size: | 12.1 Inches | Resolution: | 800×600 Pixels |
| Machine Application: | IM12 | Material: | Metal, Glass, Etc. |
| Usage: | Reusable | Condition: | Original And Used |
| Warranty: | 90 Days | Delivery Date: | 3-5 Working Days |
| Highlight: | Biocare iM12 patient monitor LCD display,LQ121S1LG55 replacement TFT screen,patient monitor maintenance parts |
||
Display for iM12 Monitor
Biocare iM12 Patient Monitor Color TFT LCD Display LQ121S1LG55 - Replacement Part for Maintenance
► Brief Description
The Biocare iM12 Patient Monitor features a 12.1-inch TFT high-resolution color display. Its primary function is to transmit and display the monitored patient physiological parameters, providing clinicians with a clear and intuitive interface for monitoring these parameters. This allows users to monitor the patient's condition at any time and make timely clinical judgments and decisions based on the monitoring data.
► Specification
|
Type |
TFT LCD Screen |
|
Dimensions |
12.1 inches |
|
Resolution |
800×600 pixels |
|
Screen Brightness |
10-level, adjustable |
|
LCD View Angle |
Horizontal / vertical view angle at least 150°/120° |
|
Model |
LQ121S1LG55 |
|
Machine Application |
iM12 |
|
Usage |
Reusable |
► Feature
► Replace the Display
1. Always disconnect the monitor's power and remove the battery to prevent short circuits. Prepare an anti-static wrist strap and specialized disassembly tools.
2. Remove the screws securing the outer casing and detach the handle. Carefully remove the side cover to expose the internal structure.
3. Locate the display screen connector, gently lift the latch, and pull out the flexible cable. Support the screen during this process to avoid damaging the fragile cable.
4. Align the new display screen cable with the connector and gently push it all the way in, then snap the latch back on. Reassemble the side cover, handle, and base in reverse order.
5. Connect the power supply and power on the monitor to check if the image display is normal.
Note: Replacement/repair operations should be performed by a professional engineer to avoid damaging the equipment.