logo
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Parts

GE Dash2000 Patient Monitor Power Supply Module 850-954M-D - OEM Replacement Part

GE Dash2000 Patient Monitor Power Supply Module 850-954M-D - OEM Replacement Part

Product Details:
Brand Name: GE
Model Number: 850-954M-D
Payment & Shipping Terms:
Minimum Order Quantity: 1 Piece
Price: USD
Packaging Details: Standard Export Packing
Delivery Time: 3-5 working days
Payment Terms: T/T,Western Union,Paypal
Contact Now
Detailed Product Description
Brand: GE Product Category: Monitoring
Product Type: Spare Parts Product Name: Power Supply Module
Model: 850-954M-D TENT 20 94V0 Machine Application: Dash 2000
Condition: Used With Good Condition Warranty: 90 Days
Delivery Date: 3 - 5 Working Days Package: Carton
Payment Method: Western Union, Bank T/T, Paypal Transportation: DHL, FedEx, UPS, TNT, EMS, By Air, Etc.
MOQ: One Piece
Highlight:

GE Dash2000 power supply module

,

patient monitor replacement part

,

OEM monitor power supply

Power Supply for Dash2000

GE Dash2000 Patient Monitor Power Supply Module 850-954M-D - OEM Replacement Part


Brief Description

The GE Dash2000, a high-quality device suitable for use in emergency rooms, intensive care units, and general wards, employs a dual power supply solution of "AC adapter power + intelligent battery system" to ensure continuity of clinical use and ease of transport. This internal power module is a crucial component of the power supply system, converting externally input unregulated DC power into multiple precision DC outputs and providing overvoltage, overcurrent protection, and undervoltage monitoring to ensure stable system operation.



► Detailed Photos

GE Dash2000 Patient Monitor Power Supply Module 850-954M-D - OEM Replacement Part 0



► More Details

The core functions of the power supply board of the GE Dash2000 monitor are AC-DC conversion, voltage isolation, and providing voltage regulation and protection. The main output voltages are as follows:

1. +5V: Powers the device's main microprocessor, digital logic circuits, memory, and other core digital components. This is the system's "heart" power supply.

2. ±5V (or ±8V): Powers the analog front-end circuits (such as amplifiers and filters for ECG, respiration, and body temperature modules), requiring low noise and high stability.

3. ±12V (or ±15V): Powers some operational amplifiers, sensors, or backlight inverters for the display screen, and other components requiring higher voltages.

4. +3.3V: Powers some low-power chips.

5. +24V: Used to drive high-power peripheral components such as printer motors or air pumps.



► Common Faults

1. Fault Symptom: The machine cannot be powered on at all; the AC indicator light is off.

Cause: Possibly a blown fuse on the power board, a damaged rectifier bridge, or a short circuit in the switching transistor.

 

2. Fault Symptom: The screen is black after powering on, but the fan spins or the sound of an internal relay engaging can be heard.

Cause: A faulty DC-DC converter circuit on the power board is preventing the generation of the high-voltage backlight voltage required for the display screen.

 

3. Fault Symptom: The machine works normally when connected to AC power, but immediately shuts down after unplugging AC power.

Cause: The internal battery is old and has failed, or the charging control circuit on the power board is damaged, preventing it from storing power.




Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Jasmine

Tel: 86-15180047496

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2026 China World Technology Medical Equipment Service Group. All Rights Reserved.