logo
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Parts

Display High Voltage Board (Inverter Board) N10134-4 for Mindray Datascope Passport2 Patient Monitor

Display High Voltage Board (Inverter Board) N10134-4 for Mindray Datascope Passport2 Patient Monitor

  • Display High Voltage Board (Inverter Board) N10134-4 for Mindray Datascope Passport2 Patient Monitor
  • Display High Voltage Board (Inverter Board) N10134-4 for Mindray Datascope Passport2 Patient Monitor
  • Display High Voltage Board (Inverter Board) N10134-4 for Mindray Datascope Passport2 Patient Monitor
  • Display High Voltage Board (Inverter Board) N10134-4 for Mindray Datascope Passport2 Patient Monitor
  • Display High Voltage Board (Inverter Board) N10134-4 for Mindray Datascope Passport2 Patient Monitor
  • Display High Voltage Board (Inverter Board) N10134-4 for Mindray Datascope Passport2 Patient Monitor
Display High Voltage Board (Inverter Board) N10134-4 for Mindray Datascope Passport2 Patient Monitor
Product Details:
Payment & Shipping Terms:
Minimum Order Quantity: 1 Piece
Price: USD
Packaging Details: Standard Export Packing
Delivery Time: 3-5 working days
Payment Terms: T/T,Western Union,Paypal
Contact Now
Detailed Product Description
Brand: Mindray Product Category: Monitoring
Product Type: Spare Parts Product Name: Inverter Board
Model: N10134-4 Machine Application: Datascope Passport2
Condition: Used With Good Condition Warranty: 90 Days
Delivery Date: 3 - 5 Working Days Package: Carton
Payment Method: Western Union, Bank T/T, Paypal Transportation: DHL, FedEx, UPS, TNT, EMS, By Air, Etc.
Size: Small MOQ: One Piece
Highlight:

Inverter Board N10134-4 High Voltage Board

,

90 Days Warranty Inverter Board

,

3-5 Working Days Delivery High Voltage Board

Mindray Datascope Passport2 High Voltage Board 

Display High Voltage Board (Inverter Board) N10134-4 with Wring Cable for Mindray Datascope Passport2 Patient Monitor


1. Brief Description

The Mindray Datascope Passport2 Patient Monitor Display High Voltage Board N10134-4, also known as an inverter or backlight driver board, is primarily responsible for illuminating the display screen so that medical staff can see the images and information on it. It has functions of boosting and driving voltage, illuminating the backlight, and controlling dimming; it is the "heart" of the backlight system. Highly integrated within the display screen, it is an indispensable component of this monitor.



► 2. Detailed Photos

Display High Voltage Board (Inverter Board) N10134-4 for Mindray Datascope Passport2 Patient Monitor 0



► 3. Common Fault 

When the high-voltage board malfunctions, the Passport2 monitor will exhibit the following typical symptoms:

1. Screen not lit, but monitor working normally:

  • You can hear the monitor starting up normally (e.g., buzzer, fan running).
  • The button backlight may be on.
  • Shining a flashlight at a certain angle on the screen sometimes reveals a very dim image (this indicates the LCD screen and motherboard are good, just lacking backlight).

2. Screen flickering or intermittent blackout:

  • The screen flickers on and off, and the symptoms may change when the device is tapped or moved. This could be due to poor soldering of components on the high-voltage board or poor contact of the high-voltage wires.

3. Uneven screen brightness or yellowing:

  • If one side of the backlight is not working or is aging, it will cause one side of the screen to be brighter than the other, or the overall brightness to decrease and the color to turn yellow.

4. Burning smell or visible arcing:

  • This is direct evidence that components on the high-voltage board (such as the high-voltage transformer, capacitors, and power transistors) have been damaged.


► 4. Notes

  • Important Warning: The high-voltage board carries a deadly high voltage during operation! Non-professional repair personnel are strictly prohibited from disassembling or inspecting it!
  • Do not replace individually: Due to its high integration, repairs typically require replacing the entire display module assembly.
  • Professional Repair: If a high-voltage board fault is diagnosed, repairs must be performed by a professional engineer.
  • Safety First: Even when the equipment is powered off, the capacitors on the high-voltage board may still store high-voltage charges, requiring a prolonged discharge period to ensure safety.



► 5. FAQ

Q1: Is this Inverter Board original?
A1: Yes, it's original from Mindray. 
Q2: Is good condition?
A2: Yes, our engineer have tested it, it's working in good condition.
Q3: How will you ship it? 
A3: We will use DHL, FedEx, UPS, TNT, EMS, By Air, etc.
Q4: How can i place the order? 
A4: We accept Western Union, Bank T/T, Paypal. Please contact us to buy it. 


Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Jasmine

Tel: 86-15180047496

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2026 China World Technology Medical Equipment Service Group. All Rights Reserved.