logo
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsDefibrillator Machine Parts

Isolated Power Supply Module With EtCO2 9301-0314-02 For ZOLL M Series Defibrillator Standalone

Isolated Power Supply Module With EtCO2 9301-0314-02 For ZOLL M Series Defibrillator Standalone

  • Isolated Power Supply Module With EtCO2 9301-0314-02 For ZOLL M Series Defibrillator Standalone
  • Isolated Power Supply Module With EtCO2 9301-0314-02 For ZOLL M Series Defibrillator Standalone
  • Isolated Power Supply Module With EtCO2 9301-0314-02 For ZOLL M Series Defibrillator Standalone
  • Isolated Power Supply Module With EtCO2 9301-0314-02 For ZOLL M Series Defibrillator Standalone
  • Isolated Power Supply Module With EtCO2 9301-0314-02 For ZOLL M Series Defibrillator Standalone
Isolated Power Supply Module With EtCO2 9301-0314-02 For ZOLL M Series Defibrillator Standalone
Product Details:
Brand Name: ZOLL
Model Number: 9301-0314-02
Payment & Shipping Terms:
Minimum Order Quantity: One piece
Packaging Details: Carton
Payment Terms: T/T, Western Union,Paypal
Contact Now
Detailed Product Description
Product Name: ZOLL M Series Defibrillator Standalone Isolated Power Supply Module With EtCO2 9301-0314-02 Brand: ZOLL
Part No: 9301-0314-02 Application: ZOLL M Series Defibrillator
MOQ: One Piece Condition: Used
Warranty: 90 Days Delivery Time: 3-5 Working Days
Payment Terms: Full Before Shipment Payment Method: T/T ,Paypal,Western Union
Highlight:

Standalone Isolated Power Supply Module

,

EtCO2 Isolated Power Supply Module

,

ZOLL Isolated Power Supply Module

ZOLL M Series Defibrillator Standalone Isolated Power Supply Module with EtCO2 9301-0314-02

 

 

► Isolated Power Supply Module

 

The isolated power supply provides an electrically isolated power supply to the EtCO2 and SpO2 modules. It also provides electrically isolated serial communication and isolated control signals between the EtCO2 and SpO2 modules and the main system PWBA.

 

► About Power Supply

 

The power supply converts DC power from a removable battery or AC/DC battery charger module to the voltage required by the M Series hardware.

The power circuit converts the raw battery or PWBA charger output voltage from +8.5 VDC to +16 VDC, including charge and line regulation, to the voltages shown in the following table.

 

 

► Power Supply Test

 

Tools needed:

 

• 1 red micro alligator test lead per micro alligator.

• 1 black micro alligator test lead per micro alligator.

• DC power supply (15 amp minimum).

• 0.1 ohm resistor 1% (¼W or more).

• 1000 ohm 1% resistor (¼W or more).

• Fluke 75 multimeter or equivalent.

 

 

Test setup:

 

Make sure the device and power supply are turned off.

 

Connect one end of the black wire to the "-" terminal in the battery compartment.

 

Connect the other end of the black wire to the "-" terminal of the power supply.

 

Connect the red wire to the "+" terminal socket of the battery compartment. Use a center peg with a protective plastic sheet around it.

 

Connect the other end of the red wire to the "+" terminal of the power supply.

 

Set the power supply voltage to 7V.

 

 

Caution

 

Be sure to connect the power supply properly to the M Series battery well terminals or damage to the unit may result. Do NOT raise the power supply voltage above 12V.

 

 

 

 

► Specification of ZOLL M Series Defibrillator Standalone Isolated Power Supply Module with EtCO2 9301-0314-02

 

Product Name

ZOLL M Series Defibrillator Standalone Isolated Power Supply Module with EtCO2 9301-0314-02

 

Brand ZOLL
Application ZOLL M Series Defibrillator
Part No 9301-0314-02
MOQ One piece
Warranty 90 days
Condition Used

 

 

 

► Photo of ZOLL M Series Defibrillator Standalone Isolated Power Supply Module with EtCO2 9301-0314-02

 

 

Isolated Power Supply Module With EtCO2 9301-0314-02 For ZOLL M Series Defibrillator Standalone 0

 

► FAQ

 

1.Will drop shipping leak supplier information?

We use neutral packaging and neutral invoices throughout the process, strictly protecting your source of goods and pricing information.

 

2.Can I specify a delivery time slot?

Some countries support scheduled delivery, while most areas follow the regular courier delivery schedule, making it impossible to specify a precise hour.

 

3.Will reshipping lost packages be expedited?

We prioritize and ship lost packages as replacements, minimizing waiting time.

 

4.Does customs clearance require the recipient's cooperation?

No cooperation is required for routine customs clearance. Customs inspection may require providing simple information, which we will provide in advance.

 

5.Is customs clearance easier with a company address?

Customs clearance is more stable with company, hospital, and physical addresses, and they are less likely to be verified than post office addresses.

 

Thank you so much for your attention .If you're interested in ZOLL M Series Defibrillator Standalone Isolated Power Supply Module with EtCO2 9301-0314-02 ,feel free to contact us .Thank you .

 

Isolated Power Supply Module With EtCO2 9301-0314-02 For ZOLL M Series Defibrillator Standalone 1 Isolated Power Supply Module With EtCO2 9301-0314-02 For ZOLL M Series Defibrillator Standalone 2
90 Day Warranty                             SGS ceryification

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Kiara

Tel: 8619854815217

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2026 China World Technology Medical Equipment Service Group. All Rights Reserved.