China World Technology Medical Equipment Service Group
Win-win together!
Brand Name: | PHiliph |
Model Number: | E116921 |
Minimum Order Quantity: | 1pcs |
---|---|
Packaging Details: | Carton |
Delivery Time: | 3-5 working days |
Payment Terms: | T/T(wire transfer), Western Union, Paypal |
Product Name: | E116921 PHiliph VS2+ Patient Monitor Power Supply Board | Brand: | PHiliph |
---|---|---|---|
Part No: | E116921 | Use For: | PHiliph VS2+ Patient Monitor |
MOQ: | One Piece | Warranty: | 90 Days |
Condition: | Used | Payment Terms: | Full Before Shipment |
Delivery Method: | DHL,UPS ,Fedex | Packaging: | Standard |
Material: | Plastic | ||
Highlight: | E116921 Patient Monitor Parts,PHiliph VS2 Patient Monitor Parts,Patient Monitor Power Supply Board |
E116921 PHiliphh VS2+ Patient Monitor Power Supply Board
PHiliphh VS2+ patient monitor Brief Introduction
The vision is believable when it comes to the enhanced SureSigns VS2+ display. With features designed to easily measure vital signs, VS2+
gives your staff the flexibility and freedom they need to spend more time with their patients. From an easy-to-read display area and large buttons
to new volume waveform displays and intuitive navigation, VS2+ gives you the information you need, when you need it. and maximize bedside
time With wireless capabilities and data export in HL7 format, the VS2+ offers reliability, peace of mind, and records patients in record time.
Power supply Board
The monitor contains a medical grade power supply module and circuitry to convert the AC line (100 V to 240 V) to DC power (15 V). Power
consumption is < 25 W.
The power management module includes:
– 10.8 V to 11.1 V
– 2200 mAH
Power management circuitry charges the Li-Ion battery through a smart battery charger. Both the battery and the AC/DC power module can
power the system through several DC/DC converters that provide the various voltage outputs required by the various chips. An isolation
transformer provides power to the front-end float area.
Power problem
Symptom | Possible cause | Action |
The display does not power up with AC power, but does with battery power. |
The power cord is unplugged | Ensure that the power cord is plugged into an outlet |
The power cord is broken | Replace the power cord if necessary | |
A fuse is blown | Replace the fuse | |
The power supply cable malfunctional | Replace the DC In cable | |
The power supply malfunctioned | Check the output voltage on the power supply .If it does not measure 15V ,replace the power supply | |
The AC power connector malfunctioned | Replace th AC connector | |
The monitor keypad is worn | Replace the front panel | |
The mainboard malfunctioned | Replace the main board |
Removing the power supply
1. Separate the front panel from the rear case assembly
2. Disconnect the power input assembly cable.
3. Remove the four Torx T10 screws and remove the power supply from the rear housing.