PHiliphh MP70 Patient Monitor Touch Screen ELO
PHiliphh Intellivue MP70 Patient Monitor Parts Touch Screen ELO Part Number: D60370-000
► Brief Description
The PHiliphh MP70 touch screen ELO part NO.D60370-000 is the part of the front display on the MP70 patient monitor, and it is a glass touch screen on the display assembly. This touch screen is a 15-inch color LCD, Up to eight waves can be shown on the screens, as well as the 12-Lead ECG Screen. On the PHiliphh MP70 touch screen, there is a cable connected to the display board, which can be used to connect to the main screen of the MP70 patient monitor for normal use.
► More Details
The PHiliphh MP70 touch screen ELO part NO.D60370-000 is a display and user interface service carrier. Applications can use advanced commands to display monitoring data or status and command windows on the internal LCD panel. These commands are interpreted by the display manager application. This application controls dedicated video hardware, which includes video memory and a dedicated ASIC. User input is obtained from various input devices such as SpeedPoint, touch screen, or other standard input devices (keyboard, mouse) that may be connected to the I/O board. System software ensures that user input is directed to the application that has the action focus.
► Integrated Touch Display not Functioning
Symptoms | Possible Causes of Failure | Failure Isolation and Remedy |
Touch Screen not functioning | Flat panel adapter not connected | Check connection from the flat panel adapter to the main board |
Touch controller not connected | Check connection from touch controller to flat panel adapter | |
Touch panel cable not connected | Check connection from touch controller board to touch panel | |
Touchscreen functionality has been permanently disabled | In service mode, select Main Setup -> User Interface and change the “Touch Enable” selection to “yes”. | |
Touchscreen functionality has been temporarily disabled | Check if touch screen functionality has been temporarily disabled (padlock symbol on Main Screen key). If yes, press and hold the Main Screen key to re-enable touchscreen operation. | |
Main board defective | Replace main board | |
Touch controller board or touch sensor defective | Replace touch controller board and touch sensor Note: Linearization data must be loaded and recalibrated after replacing the touch controller board |
|
Touch Position invalid | Touch not calibrated | Perform touch calibration : 1. Enter the Main Setup Menu 2. Select Hardware 3. Select Touch Driver Settings 4. Select Calibrate |