Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Accessories

GE Invasive Pressure IP Adapter Cable IBP Tram 451 Dual Adaptor 0.3 Meter REF 2005772-001

GE Invasive Pressure IP Adapter Cable IBP Tram 451 Dual Adaptor 0.3 Meter REF 2005772-001

  • GE Invasive Pressure IP Adapter Cable IBP Tram 451 Dual Adaptor 0.3 Meter REF 2005772-001
  • GE Invasive Pressure IP Adapter Cable IBP Tram 451 Dual Adaptor 0.3 Meter REF 2005772-001
  • GE Invasive Pressure IP Adapter Cable IBP Tram 451 Dual Adaptor 0.3 Meter REF 2005772-001
  • GE Invasive Pressure IP Adapter Cable IBP Tram 451 Dual Adaptor 0.3 Meter REF 2005772-001
GE Invasive Pressure IP Adapter Cable IBP Tram 451 Dual Adaptor 0.3 Meter REF 2005772-001
Product Details:
Brand Name: GE IP Adapter cable
Model Number: TS-F1-H2005772-001
Payment & Shipping Terms:
Minimum Order Quantity: 1
Delivery Time: 3-5 working days
Payment Terms: T/T, Western Union,PayPal
Contact Now
Detailed Product Description
Product Name: GE IP Adapter Cable Type: 2005772-001
Application: Patient Monitor Warranty: 90days
Patient Type: Adult Lead Cable Length: 30 Cm
AC Input Voltage Rate: 110V—240v Product State: Run Well
Power Supply Source: Electric Operation Policy: Quality First, Efficiency First, Reputation First, Service First
Inventory Situation: Sufficient Inventory Transport Means: Express Delivery
High Light:

Invasive Pressure IP Adapter Cable

,

2005772-001 IP Adapter Cable

,

2005772-001 Patient Monitor Accessories

 

GE Invasive Pressure IP Adapter cable IBP Tram 451 Dual Adaptor 0.3 Meter REF 2005772-001

 

 

► Brief Introduction

The dual invasive blood pressure adapter is intended to be used for connection two invasive blood pressure transducers to GE Medical System Information.

Work for GE B20 B20i B40 B105 B125 Modules. Cath-Lab, MAC-Lab, Dash 1000, Dash 2000, Dash 3000, Dash 3100, Dash 4000, Dash 5000, Eagle, Solar 7000, Solar 8000, Solar 9000, Tram 451, Tram Module 100-600, Tram Rac 3, Tram Rac 4A, Tram Scope 100-800 A, Tram Scope 100-800 L

To measure invasive blood pressure, a catheter is inserted into an artery or vein. The invasive pressure setup, consisting of a connecting tubing, a pressure transducer, an intravenous bag of normal saline, all connected together by stopcocks, is attached to the catheter. The transducer is placed at the same level with the heart, and is electrically zeroed.

The transducer is a piezo-resistive device that converts the pressure signal to a voltage. The monitor interprets the voltage signal so that pressure data and pressure waveforms can be displayed.

 

 

Invasive blood pressure measurement

Configure the simulator’s IBP channels as follows:

Sensitivity: 5 µV/V/mmHg

InvBP outputs: "0 mmHg static pressure" or "atmosphere"

  • Zeroing
  • Press IBP Zero All key.
  • Check that a message "Zeroing" followed by a message "Zeroed" is shown in the IBP parameter window.
  • Static Pressure
  • Configure the simulator's InvBP output to "200 mmHg static pressure".
  • Check that the flat pressure line appears on the related waveform field. The reading in the parameter window shall be 200 ±10 mmHg.

NOTE: Recalibration is required, if the measured value is not within the specification.

  • Pressure Waveforms
  • Configure the simulator's InvBP output to "Arterial 120/80".
  • Check that the pressure waveform for tested invasive pressure channel appears in the IBP waveform field and the Sys/Dia (Mean) pressure values are shown in the related parameter window.

 

 

► Invasive blood pressure connectors (IBP1, IBP2)

GE Invasive Pressure IP Adapter Cable IBP Tram 451 Dual Adaptor 0.3 Meter REF 2005772-001 0

1 BP_+VREF BP transducer excitation voltage, channel 1
2 BP SIG+ BP transducer signal positive (+), channel 1
3 BP_+VREF BP transducer excitation voltage, channel 2
4 AGND Analog ground
5 BP SIG+ BP transducer signal positive (+), channel 2
6 SHIELD BP cable shield
7 AGND Analog ground
8 BP SIG1 BP transducer signal negative (-), channel 1
9 BP SIG2 BP transducer signal negative (-), channel 2
10 BP1_ID BP1 probe identification
11 NC Not connected

 

 

 

If you have other inquiry of GE Invasive Pressure IP Adapter cable IBP Tram 451 Dual Adaptor 0.3 Meter REF 2005772-001, please feel free to contact us.

We are always ready to serve you.

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Rita

Tel: 8618279091952

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.