China World Technology Medical Equipment Service Group
Win-win together!
Brand Name: | Zoll Lithium Ion Rechargeable Battery |
Model Number: | 8019-0535-01 |
Minimum Order Quantity: | 1 |
---|---|
Delivery Time: | 3-5 working days |
Payment Terms: | T/T, Western Union,PayPal |
Product Name: | Zoll Lithium Ion Rechargeable Battery | Type: | Rechargeable Lithium Ion |
---|---|---|---|
Charge Time: | 4 Hours | Warranty: | 90days |
Weight: | 1.7 Lb (0.77 Kg) | Nominal Voltage: | 10.6 V |
Steady Yellow: | Battery Is Charging. | Steady Green: | Battery Is Charged. |
Alternating Yellow And Green: | No Battery Is Installed, Or There Is A Battery Charging Fault. | Patient Impedance Range:: | 25 To 300 Ω (External Defibrillation) |
Output Energy:: | 1,2,3,4,5,6,7,8,9,10,15,20,30,50,70,100,150,170,200,300,360 J | Current:: | 1.8 To 0.8 A |
Frequency:: | 50/60 Hz (±3 Hz) | Input Voltage:: | 12 VDC |
Power Consumption:: | 190 W | ||
Highlight: | 8019-0535-01,Defibrillator Lithium Ion Rechargeable Battery,Zoll R Series Defibrillator Rechargeable Battery |
Zoll R Series E Series Defibrillator Lithium Ion Rechargeable Battery 8019-0535-01 10.8V, 5.8Ah, 63Wh
► Brief Introduction
Do not operate the Zoll R Series Defibrillator without a battery. Always have a source of available backup power for all anticipated use environments. A fully charged spare battery or ready access to AC mains power from a local wall outlet can provide this backup power.Test battery packs regularly. A battery that does not pass the ZOLL charger’s capacity test might cause the R Series unit to shut down unexpectedly.When the warning LOW BATTERY appears, plug the R Series unit into a power source or install a fully charged battery pack. When the warning REPLACE BATTERY appears, immediately replace the battery pack with a fully charged pack or plug the R Series unit into a power source, as unit shut down due to a low battery condition is imminent.If mistreated, a battery pack might explode. Do not disassemble a battery pack or dispose of it in fire.
► Replacing a Battery
To remove a battery pack, press the tab on the end of the battery pack inward, and lift the battery pack out of the compartment.
To install a battery pack:
1. Place the end of the battery pack opposite the tab into the end of the compartment closest to the front of the unit.
2. Lower the tabbed end of the battery pack into the compartment and press down on the tabbed end until it locks into place.
Indicates battery status:
Battery — Verifies that the battery state of charge is sufficient for at least one hour of continuous monitoring and ten shocks at maximum energy.
► Specifications
Type | Rechargeable lithium ion |
Weight | 1.7 lb (0.77 kg) |
Nominal voltage | 10.6 V |
Recharge time | 4 hours or less within R Series |
Operating time | For a new, fully charged battery at 20°C: •100 defibrillator discharges at maximum energy (200 joules), or •4 hours of continuous ECG monitoring, or •3.5 hours of continuous ECG monitoring and pacing at 60 mA, 80 pulses per minute |
Low battery indicator | The message LOW BATTERY is displayed on the screen when there is approximately 15 minutes of ECG monitoring time left on the battery. Two-beep low battery tone sounds once a minute until just before shutdown when the unit beeps twice every 2 seconds. The time from display of the message LOW BATTERY or REPLACE BATTERY until the defibrillator shuts down varies depending on the battery age and condition |
Customer satisfaction. Our biggest pursuit! If you have any questions about Zoll R Series E Series Defibrillator Lithium Ion Rechargeable Battery 8019-0535-01 10.8V, 5.8Ah, 63Wh, please feel free to contact us at any time, we are glad to answer them.
Contact Person: Bella
Tel: 86-18879075074