Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Parts

PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118

PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118

  • PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118
  • PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118
  • PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118
  • PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118
  • PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118
  • PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118
PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118
Product Details:
Brand Name: PHiliph
Model Number: E131992 J09L078118
Payment & Shipping Terms:
Minimum Order Quantity: 1pcs
Packaging Details: Carton
Delivery Time: 3-5 working days
Payment Terms: T/T(wire transfer), Western Union, Paypal
Contact Now
Detailed Product Description
Product Name: PHiliph MP40 Patient Monitor Screen E131992 J09L078118 Brand: PHiliph
Part No: E131992 J09L078118 Condition: Used
Warranty: 90 Days MOQ: One Piece
Suitable For: MP40 Delivery Time: 2-3 Working Days
Delivery Method: DHL,Fedex ,UPS Etc Payment Method: T/T ,Paypal ,Western Union
Highlight:

Patient Monitor Refurbished Touch Screen

,

E131992 Refurbished Touch Screen

,

J09L078118 Refurbished Touch Screen

PHiliphh MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118

 

 

Touchscreen Calibration

 

To access the touchscreen calibration screen:

 

1 Enter service mode

 

2 Select Main Setup

 

3 Select Hardware

 

4 Select Touch Calibration

 

 

Perform the calibration procedure without turning off the monitor. If the screen turns off after touching the first point, the touch screen will be disabled until touch calibration is performed again.

If you calibrate the touchscreen incorrectly by selecting the wrong location, you will need to re-enter calibration mode using a different input device. If you have a support device, you can choose to reset touch calibration to the default. This creates an approximate calibration that allows you to return to the calibration menu from the touch screen.

 

 

Integrated Touch Display not functioning

 

1.Touch Screen not functioning

Possible Causes of Failure Failure Isolation and Remedy
The touchscreen function is temporarily disabled

Check if the touch screen function is temporarily disabled (lock symbol on the home button). If yes, press and hold the main button

The on-screen button reactivates the touchscreen.

The touchscreen function is permanently disabled In Service Mode, select Main Settings -> User Interface and change "Enable Touch" to "Yes".
Tablet adapter not connected Check the connection between the tablet adapter and the motherboard
The touch controller is not connected Check the connection between the touch controller and the tablet adapter
Touch screen cable not connected Check the connection between touch control panel and touch panel
Defective touch control board or touch sensor

Touch control panel and touch sensor replacement

Note: Linearization data must be uploaded and recalibrated later

Touch control panel replacement

the motherboard is broken Replace motherboard

 

2.Invalid touch position

 

Possible Causes of Failure Failure Isolation and Remedy
the touch is not calibrated

To calibrate by touch:

1. Enter the main settings menu

2. Select hardware

3. Select the touch adapter settings

4. Select Calibrate

 

PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 0

Disable/enable touch actions

There are two ways to disable/enable the touchscreen functionality:

1. Press and hold the home button to temporarily disable the touchscreen functionality. A padlock symbol appears on the key. Press and hold the home button again to reactivate the touchscreen function.

2. To permanently disable the touchscreen:

A. Enter service mode.

B. Select main settings

C. Select User Interface

D. Change Enable Touch to no.

Set Enable Touchscreen to Yes to re-enable touchscreen functionality

 

Removing the Touchscreen

 

1 Remove the front and back of the monitor as above.

2 Remove the touch controller board, press the connector safety latch, and disconnect the touch sensor cable.

Note Both the touch screen and touch controller board must be replaced for the monitor to function properly.

PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 1

Assembly Note: First connect the touch sensor cable to the touch controller board, then reinstall the touchpad.

 

3. Remove the two screws securing the power button and LED board and remove the board. Plates and screws are part of the flex panel adapter cable.

Note The screws that secure this plate are self-tapping screws. When reinstalling, first turn it in the opposite direction.

PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 2

4. Remove the 8 screws, loosen the 2 clips on the monitor frame and carefully remove the monitor assembly.

PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 3

NOTE: It is recommended that this procedure be performed in a dust-free environment. The touchscreen display assembly can be difficult to remove due to the gaskets (tape) holding the two together.

5 Remove the touch screen (sensor) from the frame.

PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 4

Note This operation can be difficult due to the seams (bands).

Assembly Note: When viewing the touch screen from the rear, the touch sensor cable must be facing to the right when placing the touch screen back into the monitor housing.

PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 5

Note Both the touch screen and touch controller board must be replaced for the monitor to function properly.

 

If you're interested in PHiliphh MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 ,feel free to contact us .Thank you .

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Kiara

Tel: 8619854815217

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2025 patients-monitor.com. All Rights Reserved.