Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Parts

Patient Monitor Printer Recorder TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series

Patient Monitor Printer Recorder TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series

  • Patient Monitor Printer Recorder TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series
  • Patient Monitor Printer Recorder TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series
  • Patient Monitor Printer Recorder TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series
  • Patient Monitor Printer Recorder TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series
  • Patient Monitor Printer Recorder TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series
Patient Monitor Printer Recorder TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series
Product Details:
Brand Name: Mindray
Model Number: TR60-F
Payment & Shipping Terms:
Minimum Order Quantity: 1pcs
Packaging Details: Carton
Delivery Time: 3-5 working days
Payment Terms: T/T(wire transfer), Western Union, Paypal
Contact Now
Detailed Product Description
Product Name: TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series Patient Monitor Printer Recorder Brand: Mindray
Model No .: TR60-F Applicated Model: IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8
MOQ: One Piece Type: Thermal Array
Speed: 25 Mm/s, 50 Mm/s Trace: 3
Warranty: 90 Days Delivery Method: DHL,UPS ,Fedex ,etc
Payment Terms: Full Before Shipment
High Light:

Patient Monitor Printer Recorder

,

Mindray IMEC7 Patient Monitor Printer Recorder

,

IPM9 Patient Monitor Printer Recorder

TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series Patient Monitor Printer Recorder

 

 

 

 

About Recorder

The recorder receives data from the motherboard and sends it to the thermal print head for printing. There is a hard button (start/stop recording) and a green LED on the front of the recorder. It is connected to the motherboard.

 

The following diagram shows its operating principle:

Patient Monitor Printer Recorder TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series 0

 

Module Description
Power interface Introduces a DC from the main board.
Power module The input power is converted to a voltage suitable for each module and then forwarded to each module.
CPU Communication between control modules.
Signal interface Controls the communication between the main board and the recorder CPU.
Motor drive circuit Receives control signals from the CPU and forwards them to the stepper engine.
Key&LED interface Includes a button and an LED which are directly controlled by the CPU.

 

 

Recorder

Type Thermal array
Speed 25 mm/s, 50 mm/s
Trace 3

Patient Monitor Printer Recorder TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series 1

 

Recorder Check

 

1. Print an ECG trace. The recorder should print correctly and the printout should be clean.

2. If the configuration recorder runs out of paper and other problems occur, the display should provide the required information. When the problem is resolved, the recorder should work normally.

 

3. Enable automatic alarm logging for all parameters and set a limit outside the alarm limit set for each parameter. In the event of a parameter alarm, the corresponding alarm recording must be activated.

 

 

 

Recorder Failures

 

  •  No printout

 

Possible Cause Corrective Action
Recorder module disabled

1. Check if the recorder status indicator is turned on

2. If yes, enable the module in the [Factory Maintenance] menu. If not, look at other possible reasons.

Paper reversed Re-install the paper roll.
Cable defective or poorly connected

1. Check that the cable between the recorder and the motherboard is properly connected.

2. Check the connecting cables and connectors for damage.

Recorder power supply failure Check if the power module outputs 5 V DC and 12V DC correctly.
Recorder failure Replace the recorder

 

 

  • Poor print quality or paper not feeding properly
Possible Cause Corrective Action
The paper roll is not installed correctly Stop the recorder and reinstall the paper roll.
Print head dirty

1. Check the thermal print head and paper roller for foreign objects.

2. Clean the thermal printhead with a suitable cleaning solution.

Print head failure Replace the print head.
Recorder failure Replace recorder.

 

 

 

Removing the Recorder (Optional)

 

1.Push the two clips aside and take out the recorder driver board, as shown below:

Patient Monitor Printer Recorder TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series 2

2.Unscrew the two PT2X6 screws and disconnect the cable and flexible cable between the recorder driver board and the recorder keyboard, as shown below. Then remove the thermal print head and recorder driver board.

Patient Monitor Printer Recorder TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series 3

3.Remove the keyboard by removing the two PT2X6 screws shown below.

 

Patient Monitor Printer Recorder TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series 4

 

If you're interested in  TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series Patient Monitor Printer Recorder ,feel free to contact us .Thank you .

 

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Kiara

Tel: 8619854815217

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.