Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsDefibrillator Machine Parts

Philip HeartStart XL M4735A Defibrillator Capacitor SP25MEG1050 B Capacity

Philip HeartStart XL M4735A Defibrillator Capacitor SP25MEG1050 B Capacity

Philip HeartStart XL M4735A Defibrillator Capacitor SP25MEG1050 B Capacity
Philip HeartStart XL M4735A Defibrillator Capacitor SP25MEG1050 B Capacity
Philip HeartStart XL M4735A Defibrillator Capacitor SP25MEG1050 B Capacity
Product Details:
Brand Name: Philip
Model Number: Heartstat XL M4735A
Payment & Shipping Terms:
Minimum Order Quantity: 1
Delivery Time: 3-5 working days
Payment Terms: T/T, Western Union,PayPal
Contact Now
Detailed Product Description
Product Name: Philip Type: Heartstat XL M4735A
Application: Defibrillator/Monitor Warranty: 90days
Product Condition: In Good Working Order KeyWords: Patient Monitor Power Supply Board
Minimum Purchase Contract: 1pcs Applications Scenes: Hospital, Clinic, Hospital Room, ICU,etc
Goodness: Original,Professional Technical Team Sale Protection: If There Are Any Problems Caused By Us, We Will Be Free For You To Repair Or Returned
Management Aims: Advanced Technology, High-quality Products, Attentive Service Inventory: Sufficient Inventory
Method Of Transport: Express Delivery AC Line Powered: 100-240 V
Humidity: Up To 95% Relative Humidity Temperature Operating: 0º C To 55ºC
Temperature Storage: -20º To 70ºC
High Light:

M4735A Defibrillator Capacitor

,

SP25MEG1050 Defibrillator Capacitor

 

philip HeartStart XL M4735A defibrillator Capacitor SP25MEG1050 B Capacity

 

 

 

Discharge the Defibrillator Capacitor

1. Lift the capacitor out of the way.

Lift up the defibrillator capacitor together with its foam base. Lift straight up, then turn the capacitor to the right and rest it on the Power PCA.

2. Inspect the Disarm Pathway.

Look for signs of arcing, burning, damage to the resistors or inductor, blown traces on the Power PCA, etc.

TIP: The Disarm Pathway is located under the Defibrillator Capacitor on the left side of the Power PCA. The Disarm Pathway consists of two large power resistors (Disarm Resistors), a red inductor, and the PCA traces that connect them.

3. Choose a discharge method.

A. If the Disarm Pathway is intact, and use the Primary Method below to discharge the defibrillator capacitor.

B. If you see any burns or damage, the Disarm Pathway may not be intact. Use the Primary Method, then also use the Secondary Method to discharge the defibrillator capacitor.

Philip HeartStart XL M4735A Defibrillator Capacitor SP25MEG1050 B Capacity 0

Primary Method

  • Using the High Voltage Discharge Tool, first touch one end of the tool’s brass bar to the current sensing resistor (the small loop of flat metal).
  • Then pivot the brass bar so its other end contacts the exposed end of the disarm resistor.
  • Maintain contact at these two points for at least 5 seconds. The defibrillator capacitor is now discharged.

 

 

 

Secondary Method

  • Using two clip leads, connect one clip lead to each end of one of the Disarm Resistors.
  • Touch the clip leads directly to the spade terminals on the Power PCA where the Defibrillator Capacitor is connected. Maintain this connection for at least 5 seconds.
  • TIP: Another way to make this connection is to clip the leads to test probe leads, then use the test probes to touch the spade terminals.
  • Repeat steps a and b using the other Disarm Resistor. The defibrillator capacitor is now discharged, even if one of the resistors is damaged.

 

► Removal

1. Lift the capacitor out of the way.

Lift up the defibrillator capacitor and its foam base. Lift straight up, then turn and rest them on the Power PCA.

2. Disconnect the capacitor.

a. Before disconnecting the capacitor, note the polarity and routing of its two wires: the red wire to the spade connector next to the red inductor, and the white wire to the spade up in the corner of the Power PCA.

b. Unplug the defibrillator capacitor from the Power PCA by pulling straight up on its spade connectors with needle nose pliers. Take care to not stress the wires.

3. Remove the capacitor.

a. Slide the wires out through the hole in the foam base.

b. Using a clip lead, short the capacitor’s terminals together to prevent charge accumulation and lay the defibrillator capacitor aside.

 

 

► Replacement

1. Replace the capacitor.

a. The new capacitor will come with a shorting bar connecting the two terminals. Disconnect the shorting bar.

b. Slide the wires through the large hole in the foam base.

2. Connect the capacitor.

Plug the terminals onto the spade connectors on the Power PCA. Orient the wires so they point toward the center of the Power PCA, as shown in Figure.

3. Restore the capacitor to its final position.

a. Pivot the capacitor around to the left.

b. Guide the white wire through the slot in the foam base.

c. Lower the foam base onto the Power PCA.

Philip HeartStart XL M4735A Defibrillator Capacitor SP25MEG1050 B Capacity 1

 

Delivered Energy Accuracy:

 

Selected Energy Delivered Energy (J) Accuracy
Load Impedance (Ohms)
25 50 100 125 150
5J 4.7 5 5.2 5.4 5.2 ±2J
10J 9.3 10 10.4 10.7 10.4 ±2J
20J 18.6 20 20.8 21.4 20.8 ±4J
30J 27.9 30 31.2 32.1 31.2 ±4J
50J 46.7 50 52.3 53.5 52.1 ±15%
70J 65.4 70 73.1 75.0 72.9 ±15%
100J 93.5 100 104.7 107.2 104.4 ±15%
150J 140.3 150 156.8 161.0 156.5 ±15%
200J 187.0 200 209.3 214.6 208.6 ±15%

 

 

 

If you have any questions about Philip HeartStart XL M4735A defibrillator Capacitor SP25MEG1050 B Capacity, please feel free to contact us at any time, we are glad to answer them. Thank you for your time! 

 

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Anna Hu

Tel: +8615083619413

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.