Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Parts

HeartStart XL M4735A Defibrillator ECG Board HiRel 220-0439-00 M4735-82100

HeartStart XL M4735A Defibrillator ECG Board HiRel 220-0439-00 M4735-82100

  • HeartStart XL M4735A Defibrillator ECG Board HiRel 220-0439-00 M4735-82100
  • HeartStart XL M4735A Defibrillator ECG Board HiRel 220-0439-00 M4735-82100
  • HeartStart XL M4735A Defibrillator ECG Board HiRel 220-0439-00 M4735-82100
  • HeartStart XL M4735A Defibrillator ECG Board HiRel 220-0439-00 M4735-82100
HeartStart XL M4735A Defibrillator ECG Board HiRel 220-0439-00 M4735-82100
Product Details:
Brand Name: Philip
Model Number: M4735A
Payment & Shipping Terms:
Minimum Order Quantity: 1
Delivery Time: 3-5 working days
Payment Terms: T/T, Western Union,PayPal
Contact Now
Detailed Product Description
Product Name: Heartstart M4735A XL Type: M4735A
PN#: 220-0439-00 M4735-82100 Warranty: 90days
Business Purpose: Advanced Technology, High-quality Products, Attentive Service Index Terms: Patient Monitor Mainboard,medicial Accessories
Strengths: Original,Technical Engineers With 12 Years Of Experience Product Features: User-friendly,Durable In Use
Stock Situation: On Sale Power Source: Electricity
Material: Plastic, PCB, Etc Power Consumption: 110V—240v
Minimum Purchase Commitments: 1 Unit Delivery Term: Express
Application: Medical Equipment,ICU,hospital Device ECG Size: 2.5, 5, 10, 20, 40 Mm/mV
AC Line Filter:: 60 Hz Or 50 Hz.
High Light:

M4735A Defibrillator ECG Board

,

220-0439-00 Defibrillator ECG Board

,

M4735-82100 Defibrillator ECG Board

 

Philip HeartStart XL M4735A defibrillator ECG Board HiRel 220-0439-00 M4735-82100

 

 

 

► Brief Introduction

Philip HeartStart XL M4735A defibrillator can be used for short-term or long-term ECG monitoring. The ECG monitoring feature allows you to:

  • Multifunctional separate electrode pads, or
  • 3 or 5-lead ECG monitoring for monitoring electrodes, configuration.

When the Heart Start XL is on, the acquired ECG will be shown on the display. The ECG can be monitored with pads, 3 leads (I, II, III) or 5 leads (I, II, III, aVR, aVL, aVF or V). The default is lead II. ECG monitoring can also display heart rate (HR) and allow you to set HR alerts.

ECG monitoring is always activated in manual mode. In AED mode, ECG monitoring is only active if the lead selection is configured on (default is on). Plug the 3-lead or 5-lead ECG patient cable into the connector labeled ECG.

 

 

 

ECG Board

The ECG Board is located in the top case, behind the printer bucket. Depending on the serial number and type of Display, your device may have a plastic shield under the ECG Board. Additionally, there are two types of plastic shields, one with a hood and one without.

 

 

 

Removal

1. Disconnect the ECG Board (four cables).

A. Disconnect the large ribbon cable by pulling straight up (connects to the Control Board ).

B. Disconnect the small 3-pin connector next to the ribbon cable (connects to the ECG Out Connector). Press on the latch to release it.

C. Disconnect the cable at the opposite end of the Board (connects to the ECG Connector). Press on the latch to release the connector.

D. Disconnect the flex circuit (connects to the SpO2 PCA). Unlatch the brown receptacle on the PCA by pulling on the top of the receptacle, then slide out the flex circuit.

2. (For devices with no shield.) Remove the ECG Board .

A. Unscrew the T10 screw located near the large ribbon connector.

B. Lift the PCA out of the case.

3. (For devices with an open plastic shield.) Remove the ECG Board .

A. Unscrew the T10 screw located near the large ribbon connector. Be careful not to break the tab.

B. Lift the PCA out of the case.

4. (For devices with a hooded plastic shield). Remove the ECG Board .

A. Unscrew the T10 screw located on the tab of the shield, near the large ribbon connector. Be careful not to break the tab.

B. Lift the PCA out of the case.

C. Remove the plastic shield, if damaged.

D. If the shield is not damaged, you can leave it in place. Otherwise, remove the two rubber standoffs and the shield.

5. Remove the ECG Board .

A. Lift the ECG Board out of the case.

HeartStart XL M4735A Defibrillator ECG Board HiRel 220-0439-00 M4735-82100 0

 

 

Frequency Response:

 

AC Line Filter: 60 Hz or 50 Hz.
Pads ECG for Display Monitor (.15-40 Hz) or EMS (1-30 Hz).
Pads ECG for Printer Monitor (.15-40 Hz) or EMS (1-30 Hz).
Leads ECG for Display Monitor (.15-40 Hz) or EMS (1-30 Hz).
Leads ECG for Printer Diagnostic (.05-150 Hz) or EMS (1-30 Hz) or Monitor (.15-40 Hz).

 

 

 

 

If you have any questions about Philip HeartStart XL M4735A defibrillator ECG Board HiRel 220-0439-00 M4735-82100, please feel free to contact us at any time, we are glad to answer them.

 

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Rita

Tel: 8618279091952

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.