Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Parts

GE Dash3000 Dash4000 Dash5000 Patient Monitor Parts Speaker Ref 419483-003

GE Dash3000 Dash4000 Dash5000 Patient Monitor Parts Speaker Ref 419483-003

  • GE Dash3000 Dash4000 Dash5000 Patient Monitor Parts Speaker Ref 419483-003
  • GE Dash3000 Dash4000 Dash5000 Patient Monitor Parts Speaker Ref 419483-003
  • GE Dash3000 Dash4000 Dash5000 Patient Monitor Parts Speaker Ref 419483-003
  • GE Dash3000 Dash4000 Dash5000 Patient Monitor Parts Speaker Ref 419483-003
GE Dash3000 Dash4000 Dash5000 Patient Monitor Parts Speaker Ref 419483-003
Product Details:
Brand Name: GE
Model Number: Dash3000 Dash4000 Dash5000
Payment & Shipping Terms:
Minimum Order Quantity: 1
Delivery Time: 3-5 working days
Payment Terms: T/T, Western Union,PayPal
Contact Now
Detailed Product Description
Product Name: Dash Type: Dash3000 Dash4000 Dash5000
Application: Medical Equipment,ICU,hospital Device Warranty: 90days
Major Material: Plastic, PCB ,etc Key Word: Speaker
Apply For: Patient Monitor Product Edge: Professional Technical Team
Stock: Sufficient Stock Product State: Working Fine
Line Voltage: 100 To 240 V Current: 1.3 To 0.5 A
Frequency: 50/60 Hz Relative Humidity: 5 To 95% At 40°C
Ambient Temperature:: 0 To 40°C (32 To 104°F) While Charging Batteries:: 0 To 35°C (32 To 95°F)
Capnostat III Sensor: 10° To 40° C (50° To 104°F) Altitude:: -610 To 4, 570 M (-2,000 To 15,000 Ft)
High Light:

Dash5000 Patient Monitor Parts

,

GE Dash3000 Patient Monitor Parts

,

419483-003 Patient Monitor Speaker

 

GE Dash3000 Dash4000 Dash5000 patient monitor speaker 419483-003

 

 

► Brief Introduction

The speaker is used for audible notification of alarms.
This internal speaker provides sound for audible alarms. For batter sound quality do not block speaker.
Audio tones are generated using a tone generator, an audio amplifier, and an 8Ω, 2.5-in. Speaker. The tone generator have built in D/A converters and a mixer to generate the dual frequency tones.
Frequencies ranging from 150 Hz to 2800 Hz are produced.

 

 

Replace speaker assembly

1. Remove the DAS assembly.

2. Remove the writer assembly/writer flex.

3. Remove the writer bracket from the frame by squeezing the sides together.GE Dash3000 Dash4000 Dash5000 Patient Monitor Parts Speaker Ref 419483-003 0

4. Disconnect the speaker cable connector from the processor board.

GE Dash3000 Dash4000 Dash5000 Patient Monitor Parts Speaker Ref 419483-003 1

5. Remove the four screw anchoring the speaker to the frame. Remove the speaker from the unit.

GE Dash3000 Dash4000 Dash5000 Patient Monitor Parts Speaker Ref 419483-003 2

6. Install the new speaker in the unit.

7. Fasten the speaker with the four screws.

8. Connect the speaker cable connector the processor board.

9. Reassemble the patient monitor in reverse order.

NOTE:

When securing the thumb screw on the writer flex, first finger tighten the thumb screw, Then, use a flat-head screw driver to tighten the screw an additional 1/4 turn.

 

 

► Speaker Test

1. Change the alarm volume of the patient monitor to 100%

2. Verify the speaker volume of the patient monitor changes accordingly.

3. Return the volume of the patient monitor to the level it was previously set to, before you changed it for this test.

 

 

► Environmental Specifications

Operating conditions:
Ambient temperature: 0 to 40°C (32 to 104°F)
While charging batteries: 0 to 35°C (32 to 95°F)
Capnostat III sensor 10° to 40° C (50° to 104°F)
Relative humidity: 5 to 95% at 40°C
Vibration: MIL-STD 810E, Method 514.4, Category 1
Altitude: -610 to 4, 570 m (-2,000 to 15,000 ft)
Storage conditions:
Maximum: 70°C (158°F) at 95% relative humidity
Minimum: –40°C (–40°F) at 15% relative humidity
CO2 Sensor: –30 to 65°C (–22 to 149°F)
Batteries: –20 to 60°C (–4 to 140°F)

 

 

 

Contact us
In order to successfully conclude the transaction, please contact us in time if you have any questions. If you want to know more information about our company or products, or if you are unclear about our warranty period, or if you have any good suggestions for our products and services, you can contact us to raise them. In short, as long as you need us, we are happy to answer you. Thank you for your cooperation. Looking forward to establishing a long-term good cooperative relationship with you.

 

 

 

All through the way we seek for providing perfect service to the satisfaction of customers. Any inquiry of GE Dash3000 Dash4000 Dash5000 patient monitor speaker 419483-003 will be highly appreciated and thanks for your support. If you have any questions, please feel free to contact us at any time, we are glad to answer them.

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Rita

Tel: 8618279091952

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.