Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsDefibrillator Machine Parts

Efficia DFM100 Defibrillator Machine Parts Patient Monitort Therapy PCA Treatment Board

Efficia DFM100 Defibrillator Machine Parts Patient Monitort Therapy PCA Treatment Board

  • Efficia DFM100 Defibrillator Machine Parts Patient Monitort Therapy PCA Treatment Board
  • Efficia DFM100 Defibrillator Machine Parts Patient Monitort Therapy PCA Treatment Board
  • Efficia DFM100 Defibrillator Machine Parts Patient Monitort Therapy PCA Treatment Board
  • Efficia DFM100 Defibrillator Machine Parts Patient Monitort Therapy PCA Treatment Board
Efficia DFM100 Defibrillator Machine Parts Patient Monitort Therapy PCA Treatment Board
Product Details:
Brand Name: Philip
Model Number: DFM100
Payment & Shipping Terms:
Minimum Order Quantity: 1
Delivery Time: 3-5 working days
Payment Terms: T/T, Western Union,PayPal
Contact Now
Detailed Product Description
Product Name: Philip Therapy PCA Treatment Board Type: DFM100
Warranty: 90days Color: Green
Key Words: Defibrillator, Material: Plastic
Usage: Outpatient Operating Room Function: Accessories For Defibrillator Machine
Advantage: Professional Engineers With 12 Years Of Experience Service: Online Technical Support
Language: English/French/Italian/Spanishand All The Other Languages Condition: In Good Working Order
High Light:

DFM100 Defibrillator Machine Parts

,

PCA Board Defibrillator Machine Parts

 

Medical Device Hospital Equipment​ Philipilip Efficia DFM100 defibrillator Patient Monitort Therapy PCA Treatment board

 

 

Brief Introduction

If the Processor PCA is the “brain” of the device, the Therapy PCA is “senses” and “muscle”. It consists of three major sections: The Power circuit, the Therapy circuit, and the Pads ECG circuit.

It performs the following functions, as directed by the Processor PCA:

  • Charges and keeps the capacitor charged to the correct energy level.
  • Delivers defibrillator shocks and controls the waveform.
  • Disarms the Capacitor and controls the disarmed energy absorption into Internal Resistors.
  • Generates pacing waveforms.
  • Charges the battery.
  • Provides analog front end for ECG from the pads/paddles.
  • Measures impedance for the Patient Contact Indicator (PCI) function.
  • Monitors the overall system power, including detecting the battery or the presence of external power.
  • Control of the RFU indicator (the RFU signal is sent through the Processor PCA).

 

Replacement

Removing the Therapy PCA involves disconnecting many cables and removing many screws. Take your time and be methodical.

 

Preparation:

  1. Open the Case.
  2. Remove the Main Chassis.
  3. Remove the Main PCA Module of the Main Chassis.
  4. Position the Therapy PCA facing you.

Removal:

  1. Disconnect the Therapy PCA connector cable, the Battery Power Cable, and the Battery Signal Cable.
  2. Remove the two M 3x15 screws with Isolation gasket.
  3. Carefully but firmly grasp the two edges of the Therapy PCA and lift the Therapy PCA.Efficia DFM100 Defibrillator Machine Parts Patient Monitort Therapy PCA Treatment Board 0

Replacement:

  1. Align the screw holes on the Therapy PCA with the screw holes on the metal plate.
  2. Secure the two M 3 x 15 screws with washers, and connect the Therapy PCA Connector Cable, Battery Signal Cable and Battery Power Cable.
  3. Assemble the Main Chassis.
  4. Install the Main Chassis.
  5. Install the Therapy Port.
  6. Install the Printer Assembly.
  7. Install the Measurement Module.
  8. Close the Case.

To Complete the Replacement:

  1. Upgrade the software to ensure the Therapy PCA firmware is compatible with the software installed on the SOM and Processor PCAs.
  2. Run Performance Verification and Safety testing.

 

Environmental Specifications

 

Storage Tenperature: -20°C to 70°C (-4°F to 158°F)
Humidity: 15% to 95% relative humidity
Operating Tenperature: 0 °C to 45°C (32°F to 113°F)
Humidity: 15% to 95% relative humidity
AtmosPhilipiliperic Pressure Range: 1060 mbar to 572 mbar (-1253 to 14,986 ft.;
-382 to 4,568 m)

 

 

Thank you for your time! If you are interested in Philipilip Efficia DFM100 defibrillator Patient Monitort Therapy PCA Treatment board, welcome to place your orders. 

Any inquiry will be highly appreciated and thanks for your support.

 

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Rita

Tel: 8618279091952

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.