logo
Tel:
Request A Quote
English
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Parts

Mindray Imec12 Patient Monitor Recorder Printer for Hospital

Mindray Imec12 Patient Monitor Recorder Printer for Hospital

  • Mindray Imec12 Patient Monitor Recorder Printer for Hospital
  • Mindray Imec12 Patient Monitor Recorder Printer for Hospital
  • Mindray Imec12 Patient Monitor Recorder Printer for Hospital
  • Mindray Imec12 Patient Monitor Recorder Printer for Hospital
Mindray Imec12 Patient Monitor Recorder Printer for Hospital
Product Details:
Brand Name: Mindray
Certification: CE
Payment & Shipping Terms:
Minimum Order Quantity: 1pcs
Packaging Details: Carton
Delivery Time: 3-5 working days
Payment Terms: T/T(wire transfer), Western Union, Paypal
Contact Now
Detailed Product Description
Product Name: Mindray Imec12 Patient Monitor Recoder Printer Brand Name: Mindray
MOQ: One Piece Warranty: 90 Days
Delivery Method: DHL,UPS ,Fedex ,etc Delivery Time: 3-5 Working Days
Application: Hospital Devices; Medical Care; ICU; Medical Equipments And Accessories Payment Method: T/T ,Paypal ,Western Union
Payment Terms: In Full Before The Shipment
Highlight:

Imec12 Patient Monitor Recorder Printer

,

Mindray Patient Monitor Recorder Printer

Mindray Imec12 patient monitor recoder printer with good condition for Hospital

 

Mindray Imec12 patient monitor recoder printer for Hospital

 

 

The recorder receives the data from the main board, and then sends it to the thermal print head for printing. There is a hard key (start/stop recording) and a green LED on the front of the recorder. It is connected to the motherboard

module Description
Power interface Introduces a DC from the main board
Power module

Convert the input power to a voltage suitable for each module, and then forward them to each module.

The figure below shows how it works

CPU Control the communications between modules
Signal interface Control the communications between the main board and the recorderCPU
Motor drive circuit Receives the control signals from the CPU and then forwards them to thestep engines
Key&LED interface Includes a button and an LED which are directly controlled by the CPU
 

 

Mindray Imec12 Patient Monitor Recorder Printer for Hospital 0

 

 

Recorder Verification Tool Required: None.

1. Print the ECG waveform. The recorder should print correctly and the printout should be clear.

2. Configure the recorder for out of paper and other issues and the monitor should be providing the appropriate information. The burner should function normally after the problem is resolved.

3.Open the automatic alarm record of each parameter, and then set the limit of each parameter outside the set alarm limit. When a parameter alarm occurs, the corresponding alarm recording should be triggered

 

 

l Recorder

To enable/disable the recorder, select [Recorder] and toggle between [On] and [Off].

CAUTION

The recorder is disabled if [Recorder] is switched off in the menu

 

 

 

 

 

l Removing the Recorder

 

 

1.Push the two clips to one side and take out the drive board of the burner, as shown in the figure below

 

 

 

 

Mindray Imec12 Patient Monitor Recorder Printer for Hospital 1

 

 

2.Unscrew the two PT2X6 screws, and disconnect the cable and flexible cord between the recorder drive board and the recorder keyboard, as shown in the figure below. Then remove the thermal print head and recorder drive board.

 

 

Mindray Imec12 Patient Monitor Recorder Printer for Hospital 2

 

 

 

 

 

 

 

3.Unscrew the two PT2X6 screws shown in the figure below and take out the keyboard

 

 

 

Mindray Imec12 Patient Monitor Recorder Printer for Hospital 3

 

l Recorder Failures

 

Symptoms Possible Cause Corrective Action
No printout

Recorder module

disabled

1.Check if the recorder status LED lights

2. If yes, enable the module in menu. Otherwise, check for other possible causes

Paper reversed Re-install the paper roll.
Cable defective orpoorly connected

1.Check that cable between recorder and main board isproperly connected

2. Check that connecting cables and connectors are notdamaged

Recorder powersupply failure Check if the power module outputs 5 V DC and 12V DCcorrectly
Recorder failure Replace the recorder.
Poor printquality orPrint head dirtypaper notfeedingproperly Paper roll notproperly installed Stop the recorder and re-install the paper roll
Print head dirty

1.Check the thermal print head and the paper roller forforeign matter.

2. Clean the thermal print head with an appropriate cleansolution

Print head failure Replace the print head.
Recorder failure Replace recorder
 

 

 

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Kiara

Tel: 8619854815217

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2025 patients-monitor.com. All Rights Reserved.