Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Accessories

P/N LI13I001A Patient Monitor Battery Mindray IMEC5 IMEC8 IMEC10 IMEC12

P/N LI13I001A Patient Monitor Battery Mindray IMEC5 IMEC8 IMEC10 IMEC12

  • P/N LI13I001A Patient Monitor Battery Mindray IMEC5 IMEC8 IMEC10 IMEC12
  • P/N LI13I001A Patient Monitor Battery Mindray IMEC5 IMEC8 IMEC10 IMEC12
  • P/N LI13I001A Patient Monitor Battery Mindray IMEC5 IMEC8 IMEC10 IMEC12
  • P/N LI13I001A Patient Monitor Battery Mindray IMEC5 IMEC8 IMEC10 IMEC12
P/N LI13I001A Patient Monitor Battery Mindray IMEC5 IMEC8 IMEC10 IMEC12
Product Details:
Brand Name: Mindray
Model Number: LI13I001A
Payment & Shipping Terms:
Minimum Order Quantity: 1pcs
Packaging Details: Carton
Delivery Time: 3-5 working days
Payment Terms: T/T(wire transfer), Western Union, Paypal
Contact Now
Detailed Product Description
Product Name: P/N LI13I001A Mindray IMEC5 IMEC8 IMEC10 IMEC12 Patient Monitor Battery New Original Brand: Mindray
Part No.: LI13I001A Battery Technology: Lithium-ions
Suitable For: Mindray IMEC Mindray IPM Voltage: 11.1V
Capacity: 5200 MhA Condition: Original New
Warranty: 90 Days
High Light:

LI13I001A Patient Monitor Battery

,

Mindray IMEC5 Patient Monitor Battery

,

IMEC12 Patient Monitor Battery

P/N LI13I001A Mindray IMEC5 IMEC8 IMEC10 IMEC12 Patient Monitor Battery New Original

 

 

Replacing a Battery

If the patient monitor is running on battery power, ensure that the patient monitor is turned off before changing the battery. To replace the battery, follow these steps:

1. Open the battery cover.

P/N LI13I001A Patient Monitor Battery Mindray IMEC5 IMEC8 IMEC10 IMEC12 0

2. Open the latch that secures the battery to be replaced and remove the battery.

3. Insert the battery into the slot with the contacts facing in.

4. Close the battery cover.

 

 

 

Description

 

Capacity 5200 mhA
Voltage 11.1V
Suitable for

Mindray iMEC

Mindray iPM

 

Part No . LI13I001A
Battery Technology Lithium-ions

 

P/N LI13I001A Patient Monitor Battery Mindray IMEC5 IMEC8 IMEC10 IMEC12 1

Condition the battery

Batteries must be conditioned before first use. The battery conditioning cycle is a continuous charge of the battery, followed by a continuous discharge and recharge of the battery. Batteries must be conditioned regularly to maintain their useful life.

 

To charge the battery, proceed as follows:

1. Disconnect the monitor from the patient and stop any observation or measurement.

2. Insert the battery to be conditioned into the battery compartment of the monitor.

3. Connect the AC / DC adapter to the monitor and let the battery charge continuously for 10 hours.

4. Disconnect the AC / DC power adapter and leave the monitor powered on by the battery until it turns off.

5. Reconnect the monitor to AC / DC power and let the battery charge continuously for 10 hours.

 

Check the battery

A battery performance test should be performed every two years before repairing the monitor or if the battery is suspected to be the cause of the problem. Rechargeable batteries can deteriorate over time. To check battery performance

please follow the instructions below:

1. Disconnect the monitor from the patient and stop any observation or measurement.

2. Connect the AC / DC power supply to the monitor and let the battery charge continuously for 10 hours.

3. Disconnect the AC / DC power adapter and leave the monitor powered on by the battery until it turns off.

4. Battery life directly reflects its performance.

If the battery life is significantly shorter than specified, replace the battery or contact support.

 

Notes

1. If the battery is fully charged and runs for too short a period of time, it may be damaged or malfunction. The operating time depends on the configuration and operation. For example, more frequent NIBP measurements also reduce operating times.

2. The battery should be replaced if it shows any visible signs of damage or if it no longer charges. Remove the old batteries from the monitor and recycle them.

 

If you're interested in P/N LI13I001A Mindray IMEC5 IMEC8 IMEC10 IMEC12 Patient Monitor Battery New Original,feel free to contact us .Thank you .

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Kiara

Tel: 8619854815217

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.