China World Technology Medical Equipment Service Group
Win-win together!
Place of Origin: | Japan |
Brand Name: | Nihon Kohden |
Certification: | CE |
Model Number: | TEC-7621K TEC-7621C |
Minimum Order Quantity: | 1pcs |
---|---|
Packaging Details: | Carton |
Delivery Time: | 3-5 working days |
Payment Terms: | T/T(wire transfer), Western Union, Paypal |
Product Name: | TEC-7621K TEC-7621C NIHON KOHDEN Cardiolife Defibrillator New Condition | Brand: | Nihon Kohden |
---|---|---|---|
Model Number: | TEC-7621K TEC-7621C | Display Size: | 5.7inch |
Segment Analyzed, Sec (ANALYSIS): | 4 | Analysis Time, Sec (ANALYSIS): | 6-12 |
Threshold, Bpm (V Tach Rate): | >180 | Threshold, MV (VF Amplitude): | >0.1 |
AC POWER: | Yes | Condition: | New & Original |
Warranty: | 90 Days | MOQ: | One Piece |
Highlight: | TEC-7621K Cardiolife Defibrillator,TEC-7621C Cardiolife Defibrillator,Nihon Kohden Cardiolife Defibrillator |
TEC-7621K TEC-7621C NIHON KOHDEN Cardiolife Defibrillator New Condition
Nihon Kohden TEC-7621K has a 5.7-inch large screen, color TFT LCD display, integrates defibrillation and monitoring.
Optional "AED" automatic external defibrillation function, and optional AED voice prompts.
Multi-parameter monitoring function, in addition to ECG and heart rate monitoring, SPO2, CO2 plug-in and NIBP plug-in can be purchased to realize multi-parameter monitoring and increase the usable range of the machine.
Defibrillation type: manual, synchronous, AED
MODEL | TEC-7621K |
Segment analyzed, sec (ANALYSIS) | 4 |
Analysis time, sec (ANALYSIS) | 6-12 |
threshold, bpm (V tach rate) | >180 |
threshold, mV (VF amplitude) | >0.1 |
AC POWER | Yes |
Output energy (50Ω in parallel):
TEC-7621C: 2, 3, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300 and 360 Joules
Energy accuracy:
2 joules: ±0.5 joules
3 joules: ±1 joule
5 to 15 joules: ±2 joules
20 to 360 joules (TEC-7621C): ±10%
Energy limitation: 50 joules of internal defibrillation
Output waveform:
TEC-7621C: Edmark, single pulse wave (parallel 50Ω)
Charging time:
TEC-7621C:
AC work: 0 to 360 joules, 5 seconds longer
AC work: 0 to 200 joules, 3 seconds
After the new battery is fully charged, when the outside temperature is 20 degrees Celsius: 0 to 270/360 Joules, 10 seconds longer
Charging display: display the charged energy value
Charging status indicator: sound after fully charged
Synchronous discharge: provided
Discharge time:
From the top of the R wave to the top of the discharge wave:
Use the ECG lead wire to detect the ECG or the electrode plate to detect the ECG: within 60ms
Use external input ECG signal: within 25ms
Large continuous cycle of charging and discharging at 270/360 Joules:
60 times: 3 times per minute, need to stop and rest for 1 minute every 1 minute of work
15 times: 3 times per minute, no rest period required
Discharge test: When 2 to 270/360 joules pass the test load, the "test normal" message will be displayed. When the selected energy fails to pass the test load, "Test Failed" is displayed.
Automatic internal discharge: The energy stored in the machine can realize automatic internal discharge under the following conditions:
1) The energy knob is turned to the discharge position
2) The power is turned off
3) The electrode plate is pulled out
4) Exceed the preset time after charging
5) The disposable electrode placed on the patient is removed
Electrode plate size:
Adult: 76X106mm
Children: 46X54mm
Length of electrode plate connection: 2.0m or longer
Electrode connection impedance:
Indicated by 3 indicator lights of different colors on the electrode plate handle
Green light: 0 to 100Ω
Yellow light: 100 to 200Ω
Orange light: greater than 100Ω
u Guardianship
Display: 5.7 inches, color LCD
Effective display range: 115X86mm
Number of stitches: 1 (ECG), or 3 (ECG, pulse*, end-tidal carbon dioxide*)
Use optional DSI access unit
Scan length: 97mm
Scanning speed: 25 or 50 mm/s (cardiograPhilipilip, pulse wave), 6.25 or 1.2550 mm/s (end tidal carbon dioxide)
Amplitude limit: 40 mm
Baseline position: fixed
Letter and number display: heart rate, blood oxygen saturation value*, pulse rate*, end-tidal carbon dioxide value*, number of breaths*, non-invasive blood pressure value** (systolic blood pressure, diastolic blood pressure, mean blood pressure)
Use optional DSI access unit
Use the optional non-invasive blood pressure unit
Other indications: synchronization, charging, discharging, fixed mode, demand mode, ~ (AC working), battery mark (indicating remaining working time), information
Alarm system:
Heart rate/pulse rate upper/lower limit (in steps of 5 beats per second)
Upper limit: 20 to 300 times per second, closed
Lower limit: 15 to 295 times per second, off
Upper/lower limit of blood oxygen saturation (in 1% steps)
Upper limit: 51 to, closed
Lower limit: 50 to 99%, closed
Pulse rate upper/lower limit (in steps of 5 beats per second)