EnviteC Oxygen Sensor OOM102-1
Technical Specifications Oxygen Sensor OOM102-1
Measurement range: | % ... 100 % oxygen (at atmospheric pressure) |
Nominal sensor lifetime: | ≥ 1 000 000 % volume oxygen hours |
Output in ambient air: | 9 mV ... 14 mV |
Electrical interface: | 3.5 mm mono phone Jack |
Accuracy: | meets ISO 80601-2-55 requirements |
Repeatability: | < 1 % volume O2 at constant temperature and pressure |
Linearity error: | < 3 % relative |
Response time: | < 12 s to 90 % of final value |
Zero offset voltage: | < 200 µV in 100 % nitrogen, applied for 5 min |
Cross interference: | meets ISO 80601-2-55 requirements |
Influence of humidity: | -0.03 % rel. per % RH at 25 °C |
Pressure range: | 0.6 bar ... 2 bar (ppO2 0 ... 1250 mbar O2) |
Influence of pressure: | proportional to change in oxygen partial pressure |
Influence of mechanical shock: | < 1 % relative after a fall from 1 m |
Operating temperature: | 0 °C ... +50 °C |
Temperature compensation: | built-in NTC compensation |
Operating humidity: | 0 % ... 99 % RH non-condensing |
Storage temperature: | -20 °C ... +50 °C |
Recommended storage: | +5 °C … +15 °C |
Recommended load: | ≥ 10 kOhms |
Warm-up time: | < 30 minutes, after replacement of sensor |
Weight: | approximately 28 grams |
Long term output drift: | < 1 % volume oxygen per month |
typically < -15 % relative over lifetime | |
Effect of temperature compensation (steady state): | between +25 °C and +40 °C: 3 % relative error |
between 0 °C and +50 °C: 8 % relative error |